OxiTop
®
Control OC 110
Operation
ba31116e05
07/2014
19
4.3
Navigation
4.3.1
Operating modes
BOD Routine
:
Measurement of single samples for a BOD
x
(x = 0.5h ... 99d) at 20 °C
The measuring range and filling volume can be selected out of seven per-
manent ranges and the corresponding filling volumes.
The function is permanently switched on
BOD Standard
:
(preselected)
Measurement of up to 12 parallel sample processes per overall sample for
a BOD
x
(x = 0.5h ... 99d) at 20 °C
The measuring range and filling volume can be selected out of seven per-
manent ranges and the corresponding filling volumes.
Automatic sample statistics with averaging
The
AutoTemp
function and the
GLP
mode can be switched on
BOD Special
:
Measurement of up to 12 parallel sample processes per overall sample for
a BOD
x
(x = 0.5h ... 99d)
Measuring range, dilution, bottle volume and sample volume can be set
The altitude above sea level and the starting air pressure respectively can
be set
The incubation temperature (5 °C to 40 °C) can be set up
Automatic sample statistics with averaging
The
AutoTemp
function and the
GLP
mode can be switched on
Pressure p
:
The
Pressure p
mode is a pure pressure measurement
The absolute pressure of single samples is measured between 500 hPa and
1350 hPa
A "limit pressure“ can be set
Up to 10 manually triggered measurements (momentary values) are saved
with pressure value and point in time
The
GLP
mode can be switched on
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......