OxiTop
®
Control OC 100
Maintenance, cleaning, disposal
ba31114e06
06/2012
61
9
Maintenance, cleaning, disposal
9.1
Maintenance
9.1.1 General maintenance activities
The only maintenance activity required is replacing the batteries.
9.1.2 Replacing the supply batteries (controller)
Switch off the controller OxiTop
®
-OC 100.
Loosen the 4 screws underneath the housing using a Phillips screwdriver.
Place the controller on the lower case.
Remove the upper case and put it down to the right next to the lower case
with the display downwards.
1 Screws
1 Supply batteries in the battery holder (lower case)
2 Data backup battery (upper case)
LISTED 8F93
Laboratory Equipment
E163694
This device complies with Part 15 of the FCC Rules. Operation
is subject to the following conditions: (1) This devise may not
cause harmful interference, and (2) this devise must accept
any interference received, including interference that may
cause undesired operation.
Ser.-Nr.:05150013
Oxi Top OC 1x0
WTW 82362 Weilheim
Made in Germany
1x CR 2430 Lithium
3x Alkaline, AA size, 1,5V
1
1
1
1
1
2