Criteria for a stable

measured value

The 

Stability control

 function checks whether the measured values are stable 

within the monitored time interval. 

The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.

7.1.2 Measuring the temperature

For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample. 
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.

7.2

FDO

®

Check procedure

7.2.1 Why should you check the sensor?

With the FDO

®

Check procedure, you can find out in a simple manner

whether the 

FDO

®

925 D.O. sensor should be cleaned or calibrated.

7.2.2 When should you check the sensor?

Checking can be useful in the following cases:

When the check interval has expired 
(the [check] status indicator is displayed). 

If the measured values seem to be implausible

If you assume that the sensor cap is contaminated or at the end of
its lifetime

After the sensor cap was exchanged

Routinely within the framework of the company quality assurance

7.2.3 Carrying out the FDO

®

Check procedure

FDO

®

Check

procedure

Check in water vapor-saturated air. 
Use the check and storage beaker (FDO

®

Check) to carry out theFDO

®

Check 

procedure.

Measured parameter

Time interval

Stability in the time 
interval

D.O. concentration

20 seconds

 : better than 0.03 mg/l

D.O. saturation

20 seconds

 : better than 0.4 %

D.O. partial pressure

20 seconds

 : Better than 0.8 mbar

Temperature

15 seconds

 : Better than 0.5 °C

Содержание Multi 9420

Страница 1: ...OPERATING MANUAL ba75964e04 04 2015 Multi 9420 DIGITAL METER FOR IDS SENSORS...

Страница 2: ...right Weilheim 2015 WTW GmbH Reproduction in whole or even in part is prohibited without the express written permission of WTW GmbH Weilheim Printed in Germany For the most recent version of the manua...

Страница 3: ...nauthorized use 11 3 Commissioning 12 3 1 Scope of delivery 12 3 2 Power supply 12 3 3 Initial commissioning 12 3 3 1 Connecting the power pack 13 3 4 General operating principles 14 3 4 1 Keypad 14 3...

Страница 4: ...n concentration 47 6 1 Measuring 47 6 1 1 Measuring the ion concentration 47 6 1 2 Measuring the temperature 49 6 2 Calibration 50 6 2 1 Why calibrate 50 6 2 2 When to calibrate 50 6 2 3 Calibration I...

Страница 5: ...measurement settings 89 9 4 D O measurement settings 92 9 4 1 Settings for D O measurement 92 9 5 Cond measurement settings 92 9 5 1 Settings for IDS conductivity sensors 92 9 6 Sensor independent se...

Страница 6: ...111 12 1 1 General maintenance activities 111 12 1 2 Exchanging the battery 111 12 2 Cleaning 112 12 3 Packing 112 12 4 Disposal 112 13 What to do if 114 13 1 pH 114 13 2 ISE 116 13 3 Dissolved oxyge...

Страница 7: ...all applications The Multi 9420 supports you in your work with the following functions proven calibration procedures automatic stability control AR automatic sensor recognition CMC continuous measure...

Страница 8: ...data from IDS sensors IDS sensors transmit the following sensor data to the meter SENSOR ID Sensor name Sensor series number Calibration data Measurement settings The calibration data are updated in...

Страница 9: ...nus automatically that do not concern this sensor To be able to use the automatic sensor recognition a meter that supports the automatic sensor recognition e g Multi 9420 and a digital IDS sensor are...

Страница 10: ...the meter Note all labels information signs and safety symbols on the meter A warning symbol triangle without text refers to safety information in this operating manual 2 1 3 Further documents provid...

Страница 11: ...The meter may only be operated according to the authorized use specified above The meter may only be supplied with power by the energy sources mentioned in this operating manual The meter may only be...

Страница 12: ...ion with the supplied power pack Operation of the system clock with a buffer battery if there is no mains power supply see section 12 1 2 EXCHANGING THE BATTERY page 111 3 3 Initial commissioning Perf...

Страница 13: ...ut voltage range of the original power pack see section 14 2 GENERAL DATA page 120 CAUTION Use original power packs only see section 14 2 GENERAL DATA page 120 1 Connect the plug of the power pack to...

Страница 14: ...or off M Selects the measured parameter CAL CAL_ Calls up the calibration procedure Displays the calibration data AR Freezes the measured value HOLD function Switches the AutoRead measurement on or of...

Страница 15: ...00 Info AutoCal e g TEC Calibration with automatic buffer recognition e g with the buffer set Tech nical buffers ConCal Calibration with any buffers Error An error occurred during calibration AR Stabi...

Страница 16: ...5 Service interface 6 Cover plate The cover plate closes the mounting slot for the IDS adapter ADA 94 IDS DIN or ADA 94 IDS BNC available as an accessory CAUTION Only connect sensors to the meter tha...

Страница 17: ...a connected sensor at any time The sensor data are available in the measured value dis play with the Info softkey 1 Channel display Display of the plug position for the respective parameter The red b...

Страница 18: ...ensors With the softkey you can very easily switch between the two display types The suitable softkey is displayed depending on the operating situation 2 Display further sensor data settings with More...

Страница 19: ...for the login to the meter If the Password menu item is grayed out no password is required for the login 1 Switch the meter on with On Off The meter performs a self test 2 Connect the sensor The mete...

Страница 20: ...e or she first logs in with a user name A valid password consists of 4 digits The user can change the password with the next login 5 Change the digit of the highlighted position with Switch to the nex...

Страница 21: ...settings and dialogs in procedures contain further subelements 5 In the Password field enter the old password with and and confirm it with ENTER 6 In the New password field enter the new password with...

Страница 22: ...ently the setting can be changed with and ENTER Example Functions Functions are designated by the name of the function They are immediately carried out by confirming with ENTER Example Display the Cal...

Страница 23: ...red value display appears The instrument is in the measuring mode 2 Open the Storage config menu with ENTER_ The instrument is in the setting mode Calibration record Calibration data storage Buffer TE...

Страница 24: ...6 Open the General submenu with ENTER 7 Open the setting mode for the Language with ENTER 8 Select the required language with General Interface Clock function Service information Reset System 01 08 20...

Страница 25: ...itched from the display of day month year dd mm yy to the display of month day year mm dd yy or mm dd yy 9 Confirm the setting with ENTER The meter switches to the measuring mode The selected language...

Страница 26: ...nge and confirm the setting with and ENTER The time is set 7 If necessary set the Date and Date format The setting is made simi larly to that of the time 8 To make further settings switch to the next...

Страница 27: ...UTOMATIC STABILITY CON TROL page 96 in the System menu The sensor connection and the USB B device interface are gal vanically isolated This facilitates interference free measurements also in the follo...

Страница 28: ...pears while the measured value is assessed as not stable A progress bar is displayed and the display of the mea sured parameter flashes As soon as a measured value meets the stability criteria it is f...

Страница 29: ...section 9 1 1 SETTINGS FOR PH MEASUREMENTS page 84 Use one to five buffer solutions of the selected buffer set in any order Below calibration with Technical buffers TEC is described When other buffer...

Страница 30: ...he AR status indicator is displayed The measured parameter flashes 8 Wait for the end of the measurement with stability control or accept the calibration value with ENTER The calibration display for t...

Страница 31: ...eted or termi nate the stability control and take over the calibration value with ENTER The calibration display for the next buffer appears voltage display 15 If necessary finish the calibration proce...

Страница 32: ...d is displayed or Switch to calibration with the next buffer with ENTER Calibration is automatically completed after the last buffer of a buf fer set has been measured Then the calibration record is d...

Страница 33: ...mperature of the buffer with 7 Start the measurement with ENTER The measured value is checked for stability stability control The AR status indicator is displayed The measured parameter flashes 8 Wait...

Страница 34: ...or in buffer solution 2 14 When measuring without temperature sensor Enter the temperature of the buffer with 15 Start the measurement with ENTER The measured value is checked for stability stability...

Страница 35: ...ity control The AR status indicator is displayed The measured parameter flashes 24 Wait for the measurement with stability control to be completed or ter minate the stability control and take over the...

Страница 36: ...ration data storage Display To open the Calibration menu press the ENTER key in the measured value display Calibration Determined val ues Displayed calibration data 1 point Asy Zero point Asy Slope Ne...

Страница 37: ...libration record to the USB B PC interface or the USB A USB printer interface Using PRT_ output all calibra tion records to the USB B PC interface or the USB A USB printer interface Quit the display w...

Страница 38: ...e measuring range Clean the IDS sensor according to the sensor operating manual Error Error 30 or 30 62 0 or 50 0 Error elimination see section 13 WHAT TO DO IF page 114 For pH IDS sensors you can opt...

Страница 39: ...for documentation 4 Measuring range for which no valid calibration is available dark gray Measured values in this range are not suitable for documenta tion Calibrate the meter with buffers covering th...

Страница 40: ...pecial QSC buffer solutions It covers the measuring range of the sensor pH 2 to pH 11 The QSC initial calibration determines the actual condi tion of the sensor and stores it as a reference in the sen...

Страница 41: ...ble to return to the sensor evaluation with the sensor symbol for this sensor 1 Open the menu for measurement settings with ENTER 2 In the QSC menu select First calibration with The calibration displa...

Страница 42: ...ion 03 04 2013 08 00 Info 1 1 Open the menu for measurement settings with ENTER 2 In the QSC menu select Control calibration with The calibration display appears AutoCal QSC Kit is displayed as the bu...

Страница 43: ...ic Stability control see section 9 6 3 AUTOMATIC STABILITY CONTROL page 96 in the System menu The sensor connection and the USB B device interface are gal vanically isolated This facilitates interfere...

Страница 44: ...e A progress bar is displayed and the display of the mea sured parameter flashes As soon as a measured value meets the stability criteria it is frozen The HOLD AR status indicator is displayed the pro...

Страница 45: ...5 Display the potential of the current zero point with CAL 6 Press ENTER to measure the reference solution The measured value is checked for stability automatic stability con trol The display of the m...

Страница 46: ...easuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement The display of the temperature indicates the active temperature measurin...

Страница 47: ...rement and calibration should not be greater that 2 K Therefore adjust the temperature of the standard and measuring solutions accordingly If the temperature difference is greater the TpErr warning ap...

Страница 48: ...is active You can terminate the Stability control function and the HOLD func tion with AR or M at any time 2 Using ENTER activate the Stability control function manually The AR status indicator appear...

Страница 49: ...and automatically switches on the temperature measurement The display of the temperature indicates the active temperature measuring mode If you wish to measure or calibrate without temperature sensor...

Страница 50: ...rocedure that uses 2 to 7 freely selectable standard solutions The concentration expected for the measurement determines the concentration of the calibration standards 1 Connect the ISE combination el...

Страница 51: ...mperature Use F4 to select the setting of the concentration 11 Set the concentration of the standard solution with and press ENTER The standard solution is measured The measured value is checked for s...

Страница 52: ...ect the setting of the concentration 16 Set the concentration of the standard solution with and press ENTER The standard solution is measured The measured value is checked for stability AutoRead 17 Wa...

Страница 53: ...storage Display To open the Calibration menu press the ENTER key in the measured value display Select the unit of the standard solution and measurement result in the ISE setup Unit menu Standard solut...

Страница 54: ...Using PRT_ output all calibra tion records to the USB B PC interface or the USB A USB printer interface Quit the display with ESC Switch directly to the measured value display with M Calibration Calib...

Страница 55: ...Multi 9420 Ser no 12345678 CALIBRATION ISE 18 01 2013 08 09 10 Sensor Ser no 12345678 Standard 1 0 010 mg l Standard 2 0 020 mg l Voltage 1 38 5 mV Voltage 2 58 0 mV Temperature 1 24 0 C Temperature 2...

Страница 56: ...of the test sample with a ther mometer 4 Open the ISE menu with ENTER 5 Thoroughly rinse the electrode with distilled water 6 Adjust the temperature of the standard solutions 7 Select Method with and...

Страница 57: ...value is achieved for the second standard calibration solution 01 08 2013 08 00 01 08 2013 08 00 1 0 010 mg l 20 2 mV 25 0 C 2 0 020 mg l 79 2 mV 25 0 C Slp 54 1 mV Sensor Ion type Ag CALIBRATION ADA...

Страница 58: ...solu tion Std conc 14 Select continue with and start the measurement with ENTER The measurement result is displayed when the measurement is com pleted 15 If necessary start measuring further samples w...

Страница 59: ...ration solutions 3 Perform a two point calibration according to user guidance 4 The calibration record is displayed as soon as a stable value is achieved for the second standard calibration solution 0...

Страница 60: ...lution Std volume and concentration of the standard solu tion Std conc 14 If necessary set the ion type for the ion in the standard Ion type with and ENTER On selection of the ion type that can be def...

Страница 61: ...9420 Ion concentration ba75964e04 04 2015 61 17 Terminate the measuring method with M A safety query appears 18 Select yes with 19 Confirm yes with ENTER Measurement with the selected method is compl...

Страница 62: ...calibration record is displayed as soon as a stable value is achieved for the second standard calibration solution 01 08 2013 08 00 01 08 2013 08 00 1 0 010 mg l 20 2 mV 25 0 C 2 0 020 mg l 79 2 mV 2...

Страница 63: ...standard solution 13 Using and ENTER enter the value for the volume of the sample Sample volume 14 Select continue with and start the measurement with ENTER The measurement result is displayed when th...

Страница 64: ...HOD page 56 2 If necessary set the ion type for the ion in the standard Ion type with and ENTER On selection of the ion type that can be defined ION Using and ENTER set the valence Valency and molar w...

Страница 65: ...d solu tion Std conc and the volume of the ISA TISAB solution ISA TISAB vol 13 Select continue with and start the measurement with ENTER When the measurement is finished an entry window appears 14 Add...

Страница 66: ...sample is calculated from the change in potential Calibration 17 If necessary start measuring further samples with ENTER Repeat steps 8 16 for all samples 18 Terminate the measuring method with M A sa...

Страница 67: ...zed water 9 Immerse the electrode in the sample that was supplemented with blank value solution 10 Using and ENTER enter the values for the volume of the sample Sample volume the volume of the ISA TIS...

Страница 68: ...ement with ENTER The measurement result is displayed when the measure ment is completed 15 If necessary start measuring further samples with ENTER Repeat steps 6 14 for all samples 16 Terminate the me...

Страница 69: ...ity of the measured medium first When the salinity correction is switched on the Sal indicator is displayed in the measuring screen 1 Connect the IDS D O sensor to the meter The D O measuring screen i...

Страница 70: ...ty control manually at any time 1 Freeze the measured value with AR The HOLD status indicator is displayed The HOLD function is active You can terminate the Stability control function and the HOLD fun...

Страница 71: ...manner whether the FDO 925 D O sensor should be cleaned or calibrated 7 2 2 When should you check the sensor Checking can be useful in the following cases When the check interval has expired the check...

Страница 72: ...r the relative D O saturation abbreviated saturation is 100 Therefore the validity scope is 98 102 The check resulted in a measured value of 99 3 1 Connect the D O sensor to the meter 2 Place the D O...

Страница 73: ...ti 9420 provides 2 calibration procedures Calibration in water vapor saturated air Use an OxiCal air calibration vessel for calibration Calibration via a comparison measurement e g Winkler titration a...

Страница 74: ...e ambient temperature 3 Start the calibration with CAL The last calibration data relative slope is displayed 4 Start the measurement with ENTER The measured value is checked for stability stability co...

Страница 75: ...indicator is displayed The measured parameter flashes 5 Wait for the end of the AutoRead measurement or accept the calibration value with ENTER The factor that was set last is displayed 6 Using set th...

Страница 76: ...ys the calibration records Further options Scroll through the calibration records with Using PRT output the dis played calibration record to the USB B PC interface or the USB A USB printer interface U...

Страница 77: ...Calibration record Error Error S 0 90 or S 1 10 Error elimination see section 13 WHAT TO DO IF page 114 Display Calibration record Relative slope CALIBRATION Ox 03 04 2013 07 43 33 FDO 925 Ser no 101...

Страница 78: ...The stability has a considerable impact on the reproduc ibility of measured values The measured parameter flashes on the display as soon as the measured value is outside the stability range when the...

Страница 79: ...not stable A progress bar is displayed and the display of the measured parameter flashes As soon as a measured value meets the stability criteria it is frozen The HOLD AR status indicator is displayed...

Страница 80: ...o ISO 7888 Linear temperature compensation lin with adjustable coefficients of 0 000 3 000 K No temperature compensation off Application tips Select the following temperature compensations given in th...

Страница 81: ...cell constant calibration in control standard You can determine the actual cell constant of the IDS conductivity sensor by calibrating with the control standard in the following range 0 450 0 500 cm...

Страница 82: ...s checked for stability stability control The AR status indicator is displayed The measured parameter flashes 6 Wait for the end of the measurement with stability control or take over the calibrated v...

Страница 83: ...s the stored calibration data to the USB A interface USB mem ory device USB printer Calibration Calibration data storage Output to RS232 USB Outputs the stored calibration data to the USB B interface...

Страница 84: ...erface USB memory device USB printer Calibration Buffer TEC ConCal NIST DIN Buffer sets to be used for pH calibration More buffers and details see section 9 1 2 BUFFER SETS FOR CALIBRATION page 86 and...

Страница 85: ...value Limit monitoring TP monitoring on off Switch on or off the limit signaling device for the tempera ture value Limit monitoring pH monitoring on pH upper limit 2 20 Upper limit A message is output...

Страница 86: ...t pH values at 1 ConCal Any Any 2 NIST DIN DIN buffers according to DIN 19266 and NIST Traceable Buffers 1 679 4 006 6 865 9 180 12 454 25 C 3 TEC WTW Technical buffers 2 000 4 010 7 000 10 011 25 C 4...

Страница 87: ...10 008 25 C 16 Metrohm 3 996 7 003 8 999 25 C 17 Beckman 4 005 7 005 10 013 25 C 18 Hamilton Duracal 4 005 7 002 10 013 25 C 19 Precisa 3 996 7 003 8 999 25 C 20 Reagecon TEC 2 000 4 010 7 000 10 000...

Страница 88: ...ORP measurement settings Settings The settings are made in the menu for measuring settings of the ORP measure ment To open the settings display the required measured parameter in the measured value di...

Страница 89: ...esets all sensor settings to the delivery condition see section 9 7 1 RESETTING THE MEASUREMENT SETTINGS page 97 Menu item Possible setting Description Calibration Calibration record Displays the cali...

Страница 90: ...cluded in the list can be measured with the setting ION Measuring with the NH 500 electrode The NH4 setting is not suitable for the gas sensitive electrode NH 500 Select the following settings Ion typ...

Страница 91: ...Multi 9420 Settings ba75964e04 04 2015 91...

Страница 92: ...possible settings are individually displayed for each sensor Below the set ting menu is displayed for two IDS sensors TetraCon 925 LR 925 01 Default settings are printed in bold Setting menu of Tetra...

Страница 93: ...Procedure for temperature compen sation see section 8 2 TEMPERATURE COMPENSATION page 80 This setting is only available for the measured parameters and Temp comp TC Linear coeff 0 000 2 000 3 000 K Co...

Страница 94: ...ng is only available for the measured parameters and Temp comp TC Linear coeff 0 000 2 000 3 000 K Coefficient of the linear temperature compensation This menu item is only available when the linear t...

Страница 95: ...ty control on off Switches on or off the automatic sta bility control during measurement see section 9 6 3 AUTOMATIC STABI LITY CONTROL page 96 System Interface Baud rate 1200 2400 4800 9600 19200 Bau...

Страница 96: ...iderable impact on the reproducibility of measured values You can activate or switch off the automatic Stability control function see section 9 6 SENSOR INDEPENDENT SETTINGS page 95 The measured param...

Страница 97: ...llowing settings for ORP measurements are reset to the default settings with the Reset function The sensor settings are reset under the Reset menu item in the menu for cali bration and measurement set...

Страница 98: ...surement settings To open the settings display the required measured parameter in the measured value display and press the ENTER key Conductivity The following settings for conductivity measurements a...

Страница 99: ...u for cali bration and measurement settings To open the settings display the required measured parameter in the measured value display and press the ENTER key Reference temperature 25 C Temperature co...

Страница 100: ...ndition The system settings are reset in the menu Storage config System Reset To open the Storage config menu press the ENTER_ key in the measured value display Setting Default settings Language Engli...

Страница 101: ...ection 10 3 1 EDITING THE MEASUREMENT DATA STORAGE page 104 and sub sequently erase the storage see section 10 3 2 ERASING THE MEASUREMENT DATA STORAGE page 105 10 2 Automatic data storage at interval...

Страница 102: ...Possible setting Explanation ID number 1 10000 ID number for the dataset series Interval 1 s 5 s 10 s 30 s 1 min 5 min 10 min 15 min 30 min 60 min Storage interval The lower limit of the storage inter...

Страница 103: ...eed as follows to switch off the automatic data storage function before the adjusted storage duration has expired 1 Remaining storage duration 2 Graphical display of the storage duration If the automa...

Страница 104: ...data storage as an example The same settings and functions are available for the automatic data storage Menu item Setting function Explanation Data storage Manual data storage Display Displays all mea...

Страница 105: ...on 10 3 1 EDITING THE MEASUREMENT DATA STORAGE page 104 10 3 3 Measurement dataset A complete dataset includes Date time Meter name series number Sensor name series number ID number Manual data storag...

Страница 106: ...e storing stable measured value Otherwise the AR dis play is missing Calibration evaluation 4 levels or no evaluation or QSC percentage 10 3 4 Storage locations The Multi 9420 meter has two measuremen...

Страница 107: ...rface USB Device Connect the Multi 9420 to the PC via the USB B interface Installation of the USB driver on the PC System requirements of the PC for installation of the USB driver PC with at least one...

Страница 108: ...rate to be selected in the range 1200 19200 Handshake RTS CTS Set at the PC only Parity none Data bits 8 Stop bits 2 1 Connect a USB memory device USB printer to the USB Host interface Model Type Pape...

Страница 109: ...ge process see section 10 2 AUTOMATIC DATA STORAGE AT INTERVALS page 101 Stored mea sured values Manual Displayed dataset with PRT after calling up from the storage All datasets with the Output to RS2...

Страница 110: ...Importer software you can record and evaluate measurement data with a PC With the aid of the BOD Analyst Pro software you can administrate BOD measurement on the PC and automatically calculate the me...

Страница 111: ...of a power failure See the relevant operating manuals of the IDS sensors for instruc tions on maintenance To retain the current setting of the date and time during the battery exchange supply the met...

Страница 112: ...e the battery from the battery compartment 4 Insert a new battery in the battery compartment 5 Close the battery compartment 1 The date day flashes on the display 6 Tighten the screws 2 of the lid of...

Страница 113: ...Multi 9420 Maintenance cleaning disposal ba75964e04 04 2015 113...

Страница 114: ...he diaphragm Extract air or moisten junction Cable broken Exchange IDS pH sensor Gel electrolyte dried out Exchange IDS pH sensor Cause Remedy IDS pH sensor The values determined for zero point and sl...

Страница 115: ...essary Temperature not stable Adjust temperature if necessary IDS pH sensor test sample Conductivity too low Use suitable IDS pH sensor Temperature too high Use suitable IDS pH sensor Organic liquids...

Страница 116: ...e electrode Electrode not suitable for the selected ion Use a suitable electrode or select a suitable ion The gas sensitive electrode NH 500 was calibrated with the NH4 setting Select the following se...

Страница 117: ...Sensor contaminated Clean the sensor Measured temperature value outside the operating conditions display of OFL UFL instead of a temperature value Keep to the temperature range for the test sample De...

Страница 118: ...by the service department Switch on the meter Open the menu ENTER_ Storage config System Ser vice information The instrument data are displayed or Connect the sensor Press softkey Info More The senso...

Страница 119: ...m the USB B interface Use a suitable USB printer see section 11 4 CONNECTING A USB MEMORY DEVICE USB PRINTER USB A INTERFACE USB HOST page 108 Check printer settings see sec tion 11 4 CONNECTING A USB...

Страница 120: ...rotective class III Test certificates CE cETLus Ambient conditions Storage 25 C 65 C Operation 0 C 40 C Admissible relative humidity Yearly mean 75 30 days year 95 Other days 85 Power supply Power pac...

Страница 121: ...ice pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipme...

Страница 122: ...x x Pseudomonas aeruginosa 99 1 Reduction After 24 Hours x x Streptococcus faecalis 99 4 Reduction After 24 Hours x x Phoma violacea 99 0 Reduction After 48 Hours x x Penicillium purpurogenum 99 3 Re...

Страница 123: ...Desc Job Ref 05B1760 x x x Test Result Unit Est Staphylococcus aureus 99 0 Reduction After 24 Hours x x x Escherichia coli 0157 99 8 Reduction After 24 Hours x x x Aspergillus niger 99 1 Reduction Af...

Страница 124: ...VERSION IS IN THE METER OR IDS SENSOR PAGE 118 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter...

Страница 125: ...date on a PC An update folder is created in the Windows start menu If an update folder already exists for the sensor or sensor type the new data are displayed there 2 In the windows start menu open th...

Страница 126: ...n ions activity divided by the unit of the molality The practical pH value is the value of a pH measurement Potentiometry Name of a measuring technique The signal depending on the mea sured parameter...

Страница 127: ...ple but the measured value is not converted TRef Meas 1 1 a T TRef D O partial pressure Pressure caused by the oxygen in a gas mixture or liquid D O saturation Short name for the relative D O saturati...

Страница 128: ...olality is the quantity in Mol of a dissolved substance in 1000 g sol vent Reset Restoring the original condition of all settings of a measuring system Resolution Smallest difference between two measu...

Страница 129: ...Firmware update 124 I Initial commissioning 12 Initialize 97 Instrument connectors 16 Interval for calibration 88 K Keys 14 M Measured value display 21 Measurement accuracy 88 Measurement data storage...

Страница 130: ...interval 101 Storing in memory 101 Automatic 101 Manual 101 T Temperature compensation 80 Temperature measurement Conductivity 80 ISE 49 O2 71 pH 28 46 Three point calibration ISE 52 pH 31 35 Transmit...

Страница 131: ......

Страница 132: ...water needs Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work We move treat analyze and return water to the environment and...

Отзывы: