Installation
MIQ/Blue PS
3 - 4
ba75573e04
05/2009
Example of determining
the main line
The following figure shows the main line L made up of the partial
lengths L1, L2 and L3 as the cable section L4 is shorter than L3:
Length of the main line = L1 + L2 + L3
Fig. 3-2
Example of determining the length of the main line
Terminator switch
Terminator switches are located in each MIQ module between the two
SENSORNET connections on the right of the terminal strip
(designation "SN TERMINATOR").
Fig. 3-3
Terminator switch
L1
= 150 m
L4 = 200 m
L2 = 200 m
L3 = 300 m
X6
X5
X4
SENSORNET 2
RED
RO
T
SHIELD
SCHIRM
GREEN
GRÜN
X3
X2
X1
SENSORNET 1
RED
RO
T
SHIELD
SCHIRM
GREEN
GRÜN
ON
EIN
OFF
AUS
SN
TERMINA
TO
R
SN Terminator