Cleaning

LR 325/001

3

Cleaning

CAUTION

Prior to cleaning, disconnect the sensor from the meter.

Outside cleaning

We recommend to clean the sensor thoroughly, especially prior to measuring 
low conductivity values. If necessary, unscrew the outside electrode from the 
shaft.

After cleaning, thoroughly rinse with deionized water and recalibrate if 
necessary.

Aging of the

conductivity

measuring cell

Normally, the conductivity measuring cell does not age. Special measuring 
media (e.g. strong acids and lyes, organic solvents) or too high temperatures 
shorten the operational lifetime considerably or damage the measuring cell. 
The warranty does not cover cases where such conditions cause failure or 
mechanical damage.

4

What to do if...

Contamination

Cleaning procedure

Lime sediments

Immerse in acetic acid for 5 minutes (volume 
share = 10 %)

Fat/oil

Clean with warm water that contains 
washing-up liquid

Error symptom

Cause

Remedy

No temperature or 
conductivity display

– Connection meter - conductivity 

measuring cell interrupted

– Cable defective

– Check connection meter - 

conductivity measuring cell 

Measurement delivers 
implausible conductivity 
values

– Cell constant incorrectly set at the 

measuring instrument

– Measuring range exceeded

– Outside electrode not completely 

screwed on

– Dirt in electrode area

– Air bubbles in electrode area

– Electrodes damaged

– Check / correct the setting

– Observe the application range

– Tighten outside electrode by 

hand as far as it will go

– Clean the conductivity measuring 

cell (see Abschnitt 3). 

– Remove air bubbles by knocking 

laterally

– Return the sensor

Incorrect temperature 
display

– The temperature sensor is not 

immersed deep enough in the 
measuring solution

– Temperature sensor defective

– Observe the minimum immersion 

depth

– Return conductivity measuring 

cell

Содержание LR 325/001

Страница 1: ...OPERATING MANUAL ba55304e06 05 2017 Flow through cell Immersion cell LR 325 001 ULTRAPURE WATER CONDUCTIVITY MEASURING CELL...

Страница 2: ...ba55304e06 05 2017 LR 325 001 2017 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...05 2017 3 1 Overview 4 1 1 Structure and function 4 1 2 Recommended fields of application 4 2 Installation 5 2 1 Changing between flow through cell and immersion cell 5 2 2 Use as flow through cell 5...

Страница 4: ...cation Measurements in ultrapure water in flow through or immersion operation depending on the mounted outside electrode 1 Connection head 2 Shaft with inside electrode and temperature sensor 3 Outsid...

Страница 5: ...hout using any tools When mounting the electrode tighten the screw connection by hand as far as it will go 2 2 Use as flow through cell To avoid the collection of air bubbles in the electrode area the...

Страница 6: ...Immerse in acetic acid for 5 minutes volume share 10 Fat oil Clean with warm water that contains washing up liquid Error symptom Cause Remedy No temperature or conductivity display Connection meter co...

Страница 7: ...25 C 77 F Dimensions in mm Weight with connection cable Immersion cell 277 g Flow through cell 487 g Materials Shaft Stainless steel 1 4571 Connection head with cable gland POM Inside electrode tempe...

Страница 8: ...0 0001 S cm 30 S cm The measuring range may be limited by the meter that is used see measuring range of your meter Temperature range 5 80 C 100 C 23 176 F 212 F Max allowed overpressure 2 x 105 Pa 2...

Страница 9: ...LR 325 001 Technical data ba55304e06 05 2017 9 Pin assignment 4 2 5 3 NTC NTC Shield not used Plug from the front 4 5 7 3 8 2 6 1 4 2 5 3 8 6 1 7 NTC 30 k Outer electrode Inner electrode...

Страница 10: ...10 ba55304e06 05 2017 Technical data LR 325 001...

Страница 11: ......

Страница 12: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: