LabStation LS Flex/430
Battery operation with LabStation LS Flex/430
5
ba75968e01
03/2012
2
Battery operation with LabStation
LS Flex/430
Battery operation
The LabStation LS Flex/430 is supplied with power with the power pack
included. If the meter and the LabStation LS Flex/430 are connected to
each other, the LabStation LS Flex/430 supplies the meter with power.
At the same time, the accumulator pack (rechargeable batteries) in the
meter is charged. The accumulator pack is charged even while the me-
ter is switched off.
Charging time of the
accumulator pack
approx. 36 hours.
CAUTION
Use original power packs only.
The line voltage at the operating site must lie within the input volt-
age range of the original power pack (
SEE
).
NOTE
The accumulator will eventually discharge even when the meter is
switched off. If you do not work with the meter for a longer period of time
you should charge the accumulator every 6 months to avoid total dis-
charging. Accumulators that are often totally discharged have a shorter
service life.
Inserting the
accumulator pack
1
2
3