Dissolved oxygen (D. O.)
147
Note
In order to accept the current measured value without
waiting for the end of the AutoRead function, simply press
g
on the terminal or measuring module or the key on the
StirrOx G probe.
Reproducibility
criteria
With identical measurement conditions, the following
applies:
Measured
parameter
Reproducibility
Response
time
D. O.
concentration
better than 0.05 mg/l
> 10 seconds
Oxygen
saturation index
better than 0.6 %
> 10 seconds
Oxygen partial
pressure
better than 0.6 mbar
> 10 seconds