Cond 7310
Conductivity
ba75933e04
06/2013
25
0.800 ... 0.880 cm
-1
(e.g. KLE 325 with a nominal cell constant of 0.840
-1
)
The cell constant is determined in the control standard, 0.01 mol/l KCl. Cell
constants outside the ranges quoted above cannot be calibrated.
In the delivery condition, the calibrated cell constant of the meter is set to 0.475
cm
-1
(conductivity measuring cell, TetraCon 325).
For this calibration procedure, the
Measuring cell
setting must be set to cal in
the
Measurement
menu. Proceed as follows to determine the cell constant:
5.2.4 Calibration data
Displaying the cali-
bration data
The calibration record of the last calibration is to be found under the menu item,
<F1>
/ [Menu] /
Calibration
/
Calibration record
. To open it, press the
<CAL_>
key in the measured value display.
1.
Connect the conductivity measuring cell to the measuring instrument.
2.
In the measured value display, select the conductivity parameter with
<M>
.
3.
Start the calibration with
<CAL>
.
The cell constant that was calibrated last is displayed.
4.
Immerse the conductivity measuring cell in the control standard solu-
tion, 0.01 mol/l KCI.
5.
Start the measurement with
<ENTER>
.
The [AR] status indicator is displayed. A progress bar is displayed and
the display of the measured parameter flashes.
6.
Wait for the end of the measurement with stability control or accept the
calibration value with
<ENTER>
.
The calibration record is displayed and output to the interface.
7.
Using
<F1>
/
[Continue]
or
<ENTER>
, switch to the measured value
display.
The calibration record is automatically transmitted to the interface
after calibrating.
ϰ
0.479 1
/cm
25.1
°C
[Tr25][nLF][ 0.479 1/cm]
01.07.2011
08:00