Multi 340i
Resetting to default settings
13
Resetting to default settings
You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
Measurement
parameters
The following measurement parameters (pH/O
2
/
InI) can be reset to the de-
livery status:
Configuration
parameters
The following configuration parameters (InI) can be reset to the delivery sta-
tus:
Procedure of the reset:
pH/ORP (pH InI)
Measuring mode
pH
Asymmetry
0 mV
Slope -59.16
mV
Oxygen (O
2
InI)
Measuring mode
D. O. concentration
Relative slope
1.00
Salinity (value)
0.0
Salinity (function)
off
Conductivity (
InI)
Measuring mode
Cell constant
0.475 cm
-1
Baud rate
4800
Display during pH calibration
Buffer nominal value
Interval 1
(automatically saved)
OFF
Interval 2
(for data transmission)
OFF
l
To switch it off, press <CAL> while pressing the <RUN/ENTER> key.
l
To reset the measurement parameters with <
▲
> <
▼
>, select YES and
confirm with <RUN/ENTER>, or
l
continue with the next parameters without a reset with the <RUN/EN-
TER> key (O
2
InI >
InI > InI).
l
After the configuration parameters, InI changes the instrument to the last
active measurement mode.
o
n
n i
i
pH