wtw Br 500 BNC Скачать руководство пользователя страница 7

CN 500 / CN 800

Aging

ba15356e03

09/2014

7

Aging

Please note that every electrode undergoes a natural aging process.The re-
sponse time increases and the slope decreases with the age of the electrode. 
The following factors shorten the lifetime considerably:

Unsuitable storing conditions (see section S

TORAGE

)

Special measuring conditions (e.g. organic solutions, frequent measure-
ment with high concentrations of interfering ions)

High temperatures

High changes in temperature

The warranty does not cover damage caused by measuring conditions and 
mechanical damage.

Maintenance

Combination electrodes:
Refill any used up bridge electrolyte.

In the case of increased response times of the electrode perform the fol-
lowing activities:

– Polish the membrane surface using an SF/K polishing strip (see section 

R

ECOMMENDED

 

ACCESSORIES

). To do so, moisten a polishing strip with 

deionized water, and with circular movements polish the membrane 
surface for approx. 30 seconds.

– Then recondition the electrode (see section C

ONDITIONING

CALIBRATION

MEASUREMENT

).

Clean the inside of the combination electrode. To do so, open the 
combination electrode as follows:

Opening the

combination

electrode

NOTE

Never tear at the connection cable of the electrode.
The cable might be damaged.

1

Unscrew the knurled nut from the connection head.

2

Position the connection cable upward in a straight line with the con-
nection head.

3

Unscrew the connection head.

4

Push the connection head and pressure spring over the connection 
cable.

5

Push the inner body out of the shaft.

Содержание Br 500 BNC

Страница 1: ...OPERATING MANUAL ba15356e03 09 2014 CN 500 DIN CN 500 BNC CN 800 DIN CN 800 BNC CYANIDE SENSITIVE ELECTRODE ...

Страница 2: ... of the present operating manual can be found on the Inter net under www WTW com It is also available in other languages Copyright 2014 WTW GmbH Reprinting even as excerpts is only allowed with the explicit written authorization of WTW GmbH Printed in Germany ...

Страница 3: ...iew 1 Connection head with connection cable 2 Shaft 3 Membrane 4 Knurled nut 5 O ring 6 Filling opening for bridge electrolyte 7 Inner junctions 8 Ground junction 9 Storage vessel for CN 800 10 Protective cap for CN 500 CN 800 CN 500 7 9 7 8 4 6 5 3 2 1 10 ...

Страница 4: ...electrode is re quired e g R 503 The two electrodes together form a double rod combina tion electrode 1 Remove the storage vessel 2 Fill the bridge electrolyte ELY BR 503 into the filling opening 3 Shortly press the shaft of the combination electrode against the con nection head to wet the ground junction with bridge electrolyte 4 When doing so refill any spilled bridge electrolyte so that the inn...

Страница 5: ...rode is sufficiently conditioned when the measured value no longer changes considerably under otherwise constant conditions temperature composition of the solution NOTE Avoid measuring samples with cyanide concentrations of more than Minimum depth of im mersion The ground junction 4 must be covered Maximum depth of im mersion Approx 1 cm below the fluid level 2 of the bridge electrolyte 1 2 3 4 CN...

Страница 6: ...uring It provides a constant ionic strength and similar diffusion potentials at the reference electrode in standard solution and test sample If you would like to have more detailed information concerning sample prep aration and measuring procedures WTW provides a large number of appli cation reports for various applications Response times The response time depends on the concentration range It is ...

Страница 7: ...nse times of the electrode perform the fol lowing activities Polish the membrane surface using an SF K polishing strip see section RECOMMENDED ACCESSORIES To do so moisten a polishing strip with deionized water and with circular movements polish the membrane surface for approx 30 seconds Then recondition the electrode see section CONDITIONING CALIBRATION MEASUREMENT Clean the inside of the combina...

Страница 8: ...rode in a dry place CN 800 Remove the bridge electrolyte then rinse the combination electrode with deionized water and dab it dry with a clean paper towel Store the combi nation electrode in the storage vessel in a dry place For the following measurement prepare the combination electrode as described in section COMMISSIONING Store the reference electrode according to the instructions in its operat...

Страница 9: ...ve electrode ba15356e03 09 2014 9 Calibration line of a cyanide sensitive electrode 10 fache Konzentrationsänderung Meßkettenspannung 10 fold concentration change Electromotive force 1 0 1 10 100 1000 58 mV mg l CN 10 10 10 10 10 mol l CN 10 5 6 3 2 1 4 ...

Страница 10: ... for some hours until the crusts have dissolved Ground junction contaminated CN 800 Rinse ground junction with bridge electrolyte Cable broken Exchange the electrode Cause Remedy Membrane surface contaminated Polish the membrane surface with polishing strip S FK see MAINTENANCE Conditioning time too short Extend conditioning time Standard solutions too old Use new standard solutions Inner junction...

Страница 11: ...ug or BNC plug depending on design 0 2 260 mg l CN recommended 0 2 25 mg l CN 8 x 10 6 1 x 10 2 mol l CN recommended 8 x 10 6 1 x 10 3 mol l CN 2 0 14 Before measuring adjust the pH value of the sample to pH 10 to 14 0 80 C shortly up to 100 C 30 M CN 500 170 mm including 50 mm connection head CN 800 153 mm including 33 mm connection head Shaft 12 mm Connection head 16 mm 1 m ...

Страница 12: ...12 ba15356e03 09 2014 Technical data CN 500 CN 800 ...