Manual do Usuário | User Guide – WTA
26
To enable this function simultaneously press the "+" (01) and "menu" (02) keys for 3
seconds.
To exit the economy mode, simultaneously press the "-" (01) and "menu" (02) keys for 3
seconds.
Activation/deactivation of this mode can also be done by the application via Bluetooth.
Note:
During power saving mode, the "ON/OFF" button (03) should be in the "ON" position.
If the equipment is switched off during economical mode, the unit will be started
with the "economy mode" function is deactivated (display and audible alarm will
be active) and the temperature will be resumed at the same value previously set.
Note:
For greater autonomy, perform the first heating using the external power supply.
Battery Charge Warning
When charge battery approximate to 0%, the equipment beeps to indicate low charge,
this warning will sound for about 5 minutes after that period the equipment will be
deactivated to prevent the damage battery
Note:
To ensure the battery life, avoid reaching a critical level of 0%.
If you will not be using the equipment for a long time, unplug the power cable.
OOCYTES MICROTUBES ACCOMMODATION
Once the temperature is stabilized the microtubes containing the oocytes can be
inserted into the wells. Initially place the handles in the peripheral wells of the storage block.
The 4 central wells have a built-in tube lifter and do not need handles. Finally, make sure
that the cover (07) is properly closed.
ALARMS
The equipment will emit intermittent audible alarms when the storage block temperature
varies ± 2.0° F or when the battery charge level reaches close to 0%.