2
World Precision Instruments
SPECIFICATIONS
ELECTRICAL:
115 VAC, 60 Hz
PRIMING:
Unit is self priming and will hold vacuum when off.
OPERATION:
Pump can run dry without damaging the motor or drive unit, however
the tubing life will be reduced.
CONSTRUCTION:
High impact plastic and long lasting alloy metals housed in a steel
enclosure.
MAX. HEAD PRESSURE:
40 PSI. Normal pressure 18 PSI varies depending on tubing
used.
MAX. SUCTION LIFT:
29 feet (8.8m) with H
2
O.
RATING:
100% continuous operation at maximum of 40°C ambient.
FLOW RATE:
Variable speed drive system - reversible.
FUSE:
Located in the inlet receptacle. Uses 5 X 20mm - 1.0 amp. (See page 4 for
replacement procedure.)
DC MOTOR BRUSHES:
Replaceable (See page 4
for replacement procedure.)
WEIGHT:
13 lb
Included with MityFlex are four short lengths of Norprene
™
tubing to fit into the roller
mechanism —
1
/
16
”,
1
/
8
”,
3
/
16
”,
1
/
4
” — and black rollers for use with
3
/
16
” to
1
/
4
” tubing. The
optional
MITY-KIT
includes replacement tubing and a set of red rollers for use with
1
/
16
” and
1
/
8
” tubing.
FLOW RANGE:
Tubing ID
Flow (mL/min)
1
/
16
” (0.062”/1.6mm)
1.7 - 48.0
1
/
8
” (0.125”/3.1mm)
7.0 - 190.0
3
/
16
” (0.187”/4.8mm)
16.0 - 456.0
1
/
4
” (0.250”/6.4mm)
28.0 - 800.0
Flow range based on pumping water at 25°C and 0 PSI using 60 durometer
thermoplastic elastomer tubing.