background image

ISO-DAM

supply. With the Power switch in the OFF/CHArGE position, the red lamp above the 
switch will light while the batteries are charging. When depleted, batteries can usually 
be recharged in 12 to 14 hours. They should not be damaged by charging for longer 
periods.

Power Adapter

The wall power adapter (power supply) converts mains alternating current to 12 volts DC 
to power ISO-DAM. Note that the power receptacle on top of the unit is dual. Thus, using 
a jumper cable (provided), up to four ISO-DAM amplifiers can be operated from one 
power adapter. Jumper cables are supplied for each amplifier so that ISO-DAMs can be 
multiply connected to 12 volt DC power. The user may also substitute a 12 volt storage 
battery as a power source rather than line power if desired.

Specifications

Voltage Gain 

×

10, 

×

100, 

×

1000, 

×

10,000

Input r to Gnd, each input 

> 10

12

 Ohms

Input r, between Inputs 

> 10

12

 Ohms

Amp Noise, input shorted 

< 0.4  µV rms (1 Hz-100 Hz), 

 

 

 

 

< 2 µV rms (1 Hz-10 kHz)

Bandwidth 

DC to 10 kHz 

 

 

 

 

 

Filter, low pass 

0.1, 0.2, 0.5, 1, 3, 10 kHz   

 

 

 

Filter, high pass 

0.1, 1, 10, 300 Hz

Input Offset V. Adjust. 

± 100 mV

Common Mode rejection 

> 100 dB @ 50/60 Hz

Output Voltage Swing 

± 5 Volts

Output resistance 

220 

Ω

Power Adapter 

115 V 60 Hz or    

 

 

 

 

 

230 V 50 Hz to 12 V DC (user specifies)

Dimensions 

×

 4 

×

 1.75 in. (20.3 

×

 10.2 

×

 4.4 cm)

Weight 

1.5 lb (0.7 kg) approx.

Shipping Weight  

3.5 lb (1.6 kg)

WOrLD PrECISION INSTruMENTS

Introduction

The ISO-DAM preamplifier is photoelectrically isolated from earth ground. No significant 
current can flow from input terminals of the instrument to ground. The instrument is 
intrinsically safe and cannot cause any electrical stimulus or shock to the subject. ISO-
DAM’s small case allows the instrument to be placed close to the site of measurement. 
ISO-DAMs can be stacked in groups of four in a convenient 19-in. rack mount.

The input stage of ISO-DAM is battery powered by two long-lived miniature Ni-Cd 
batteries and has no electrical connection to the AC power circuit ground. This isolation 
is achieved by photoelectric coupling. Isolated preamplification has several important 
advantages over non-isolated amplifiers. 

WOrLD PrECISION INSTruMENTS 

1

Faraday ground

BLACK

+

RED

BRAID

(Ground Reference)

Содержание ISO-DAM

Страница 1: ...venage Hertfordshire SG2 7EG Tel 01438 880025 Fax 01438 880026 E mail wpiuk wpi europe com Germany Liegnitzer Str 15 D 10999 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com Internet www wpiinc com www wpi medical com www nitricoxide net www pipetter com ISO DAM Isolated low noise preamplifier World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 031507 w...

Страница 2: ...ISO DAM ...

Страница 3: ...his publication may be reproduced or translated into any language in any form without prior written permission of World Precision Instruments Inc Contents Introduction 1 Noise and Interference 2 Quick Test 2 Input Connection 2 DC Bal 3 Output 3 Filters 3 DC DC Offset 3 Charge 3 Power Adapter 4 Specifications 4 Warranty 5 ...

Страница 4: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or indirectly by use of this product Claims and Returns Inspect all shipments upon receipt Missing cartons or obvious damage to cartons should be noted on the delivery receipt before signing Concealed l...

Страница 5: ...ss 0 1 0 2 0 5 1 3 10 kHz Filter high pass 0 1 1 10 300 Hz Input Offset V Adjust 100 mV Common Mode Rejection 100 dB 50 60 Hz Output Voltage Swing 5 Volts Output Resistance 220 Ω Power Adapter 115 V 60 Hz or 230 V 50 Hz to 12 V DC user specifies Dimensions 8 4 1 75 in 20 3 10 2 4 4 cm Weight 1 5 lb 0 7 kg approx Shipping Weight 3 5 lb 1 6 kg 4 World Precision Instruments Introduction The ISO DAM p...

Страница 6: ... input socket is a 3 mm banana type socket one banana plug is also included with the instrument The user should connect this receptacle to the system ground point i e that point where the recording site is connected to Power Mains ground The subject or 2 World Precision Instruments preparation should also be connected to ground Input protective devices limit the input differential signal from exce...

Отзывы: