4. GENERAL DEVICE INFORMATION
In both versions, there is an RS485 bus connector. You can use the same commands as on USB 4.1.2. BAUDRATE
is 9600.
4.1.2 USB Communica on
A USB-C cable can be used to set the parameters of the Workswell BB200 on versions 1.4.17 and above. Connect
the device to the computer with a USB-C cable. The device must be connected to the power supply at all mes,
connec ng only to the USB-C without supplying power to the device will not suffice and you will not be able to
communicate with the device. A er connec ng to the PC and installing the driver, the device will show up as a
virtual COM port named USB-SERIAL CH340. Set this port in Pu y, CMD or other. Set BAUDRATE to 115200. In
the serial line se ngs, set 8 data bits, 1 stop bit, and no parity bit (8N1).
4.1.3 Command protocol
Communica on using ASCII characters.
List of commands for communica on with the device:
SET C/F <NNN.N>
set setpoint to <NNN.N> °C/°F temperature
SET C/F ?
shows current setpoint
GET C/F ?
shows current temperature
GET IN C/F ?
shows current temperature of the electronics
DISP C/F
switches the temperature shown on the display to °C or °F
LOCK ?
locks the temperature
UNLOCK ?
unlocks the temperature
ARTICLE ?
shows ar cle number
SERIAL ?
shows serial number
VERSION ?
shows FW version number
BUZZER ON
turns on the buzzer
BUZZER OFF
turns of the buzzer
LED <NNN>
sets the LED intensity as a percentage of 0 -100 %, where 0 does
not light up the LED at all
LED ?
shows current LED intensity
Note:
Commands listed above can be used with °C or °F. For each C/F command choose one of the op ons.
Example:
SET C ?
shows current temperature in °C,
SET F ?
shows current temperature in °F.
E-mail and Web
[email protected]
www.workswell.eu
Mobile:
+420 725 877 063
ID:
Reg. No.: 29048575
VAT No.: CZ29048575
Headquarters
U Albrechtova vrchu 12
155 00, Prague 13
Czech Republic
Revision 1.6.2 EN, 7