89 | UM-PP-EN-03
8.6
Interface Wire
To control the treadmill via PC/notebook, the serial interface of the treadmill must be connected with
the computer via a null-modem cable with 2 x D-SUB connectors (9-pin, female). This interface wire
must be shielded and must not exceed a length of 16.5 ft. (5 m). The cable is commercially available
or can be obtained from WOODWAY Customer Service.
Fig. 52 USB interface wire RS-232
There are null-modem cables with different configurations (depending on whether a hardware
handshake and/or hardware flow control is desired). A typical configuration is shown in the following
table. For controlling a WOODWAY treadmill usually only the signals GND, TxD, and RxD are required.
If a minimal null-modem cable is used, within the respective connector the pins 1, 4, 6, as well as 7
and 8 should be connected (see Fig. 52).
Plug 1
(D-SUB, 9 pin) Pin (Function)
Plug 2
(D-SUB, 9 pin) Pin
1 (DCD)
7 + 8
2 (RxD)
3
3 (TxD)
2
4 (DTR)
6
5 (GND)
5
6 (DSR)
4
7 + 8 (RTS + CTS)
1
Parameters
The WOODWAY standard software uses the following interface parameters:
•
9600 Bd (baud)
•
8 bits
•
No parity
•
1 stop bit
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...