ULTRASONIC SCALER INSTRUCTION MANUAL

Содержание U6 LED

Страница 1: ...U6 LED ULTRASONIC SCALER INSTRUCTION MANUAL...

Страница 2: ...justment 5 3 Maintenance 6 4 Cleaning Disinfection and Sterilization 8 5 Precaution 13 6 After service 15 7 Symbol instruction 15 8 Environmental protection 16 9 Manufacturer s right 16 10 European au...

Страница 3: ...mance and structural composition Ultrasonic scaler U6 LED is composed of electrocircuit water way and ultrasonic transducer 1 2 3 Scope of application Ultrasonic scaler U6 LED is used for the dental c...

Страница 4: ...trous Oxide Equipment is not suitable for being used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 3 17 The main unit use a power adapter to connet network power...

Страница 5: ...3 1 4 2 Back of the main unit sketch map picture 2 1 4 3 Instruction of using wrench Picture 3...

Страница 6: ...4 Picture 4 1 4 4 List of removable parts Picture 5 Picture 6...

Страница 7: ...ure 4 2 1 6 Turn on the power switch If the power indicator is lighted the machine is ready for work 2 1 7 Under normal working condition the frequency of the tips is very high Light touch and a certa...

Страница 8: ...mething wrong with the handpiece Send the handpiece to our company to repair Something wrong with the cable Contact our dealers or us The scaling tip vibrates but there is no spray when stepping on th...

Страница 9: ...from the endochuck The screw is loose Tighten it LED light don t work Poor contact Contact tightly Something wrong with LED light Change a new one There is no water coming out from the handpiece auto...

Страница 10: ...aged a The vibrating intensity and the pulverization degree become weak obviously b During operating there is some buzz when the scaling tip is working Note 4 If the power cable connected to the grid...

Страница 11: ...ons 4 1 2 Post operative treatment The post operative treatment must be carried out immediately no later than 30 minutes after the completion of the operation The steps are as follows 1 Let the Ultras...

Страница 12: ...th the product b In washing stage the water temperature should not exceed 45 C otherwise the protein will solidify and it would be difficult to remove c After cleaning the chemical residue should be l...

Страница 13: ...Direct use after disinfection temperature 90 C time 5 min or A0 3000 d2 Sterilize it after disinfection and use temperature 90 C time 1 min or A0 600 d3 For the disinfection here the temperature is 9...

Страница 14: ...tact with parts of different metals when packaging 4 8 Sterilization Use only the following steam sterilization procedures fractional pre vacuum procedure for sterilization and other sterilization pro...

Страница 15: ...fection of main unit are as follows Before each use wipe the surface of the main unit the connector of cable and cable with a soft cloth or paper towel soaked in 75 medical alcohol Repeat the wipe for...

Страница 16: ...n while the device is used for treatment 5 1 16 The screw thread of the scaling tips produced by other manufacturers maybe coarse rusty and collapsed which will damage the screw thread of the handpiec...

Страница 17: ...year free repair by providing warranty card The repair of the equipment should be carried out by professional technician We are not responsible for any irretrievable damage caused by non professional...

Страница 18: ...d representative 11 EMC Declaration of conformity The device has been tested and homologated in accordance with EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be eff...

Страница 19: ...ines 2kV for power supply lines 1kV for interconnecting cable Mains power quality should be that of a typical commercial or hospital environment Surge lEC 61000 4 5 1 kV line to line 2 kV line to eart...

Страница 20: ...ey should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the following symbol NOTE I At 80 MHz end 800 MHz the higher frequ...

Страница 21: ...at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum...

Страница 22: ...ZMN SM 128 V1 4 2020317...