3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
87
3.7 Sheath Flow Syringe Assembly Replacement
3.7.1 Replacing the syringe
Purpose
The syringe can be disassembled and replaced by following the procedures specified in this section;
however, in each case,
3.2 Preparatory Work before Repairs
need to be taken before you begin.
Tools/Spare Parts
⚫
2.5mm hex-socket screwdriver
⚫
The replacement syringe assembly with the same specifications
Preliminary Steps
⚫
Open the left side door.
Disassembly
1. Pull out the peripheral fluid tubes connected to the syringe assembly to be replaced.
2. Use the 2.5mm hex-socket screwdriver to disassemble the two M3×8 stainless-steel inner hex
screws affixing the syringe assembly. Then take out the syringe assembly. See Figure 3-19.
Figure 3-19 Replacement of syringe
1
2
1- Syringe module
2- M3×8 stainless-steel inner hex screw (×2)
Installation
Follow the corresponding disassembly steps in reverse order.