2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
30
Figure 2-18 Cleaning the swab
4. After cleaning and reassembling the swab, check and make sure the inflow/outflow interface
tubing is connected correctly.
5. Turn on the device.
6. Go to
Service
>
Maintenance
>
Clean
. Click the
Clean Sample Probe
icon to perform the
operation before performing cleaning and maintenance on the swab.
2.2.11.2 Cleaning and Maintenance of the WBC Bath
Go to
Service
>
Maintenance
>
Clean
and click the
Clean
WBC bath
button. See Figure 2-19.
Figure 2-19 Cleaning and maintenance interface
2.2.11.3 Cleaning and Maintenance of the RBC Bath
Go to
Service
>
Maintenance
>
Clean
and click the
Clean
RBC bath
button.
2.2.12 Troubleshooting Common Hydraulics Problems
2.2.12.1 Commonly Used Equipment and Tools
Name
Illustration
Usage
Remarks
small sample cup
⚫
Used to contain liquid after a tube is
pulled out
⚫
Used as a diluent container
⚫
The ratio of probe cleanser to
diluent is 1:3.
None