Specification

Nominal speed: 

0-11000 RPM

*

Maximum G-force:

8,200 g

Capacity: 

4 capillary tubes

Max. Power: 

12 volt DC

Height: 

11.4cm (4.5 inches)

Width: 

15cm (6 inches)

Depth: 

20cm (8 inches)

Weight:

1.58kg (3.5 lb.)

Boxed dimensions:

21.59 x 26.67 x 15cm
(8.5”) x (10.5”) x (6”) in

Boxed weight:

2.73kg (6 lbs.)

Average Speed:*   These are average speeds, which may vary
slightly between units and with load sizes. Please use a
tachometer for actual RPMs on your unit and for calibration
purposes.

Warranty

Your Clinispin Micro centrifuge is covered by a 
limited 1-year warranty on parts and labour. If the
centrifuge or any of its parts should fail during this
period, simply contact Woodley Equipment Company
Ltd.  Any attempts to repair the centrifuge or replace
any of its parts without authorisation from Woodley
Equipment Company Ltd may void the warranty. 

Instructions for use

1

Spin only balanced loads. Make sure that capillary tubes are 
placed opposite each other.

2

Insert capillary tube(s) and close lid.

3

Set timer for 5 minutes - the unit will immediately begin spinning 
at full speed (11,000 rpm).

4

Do not open the lid at any time whilst the centrifuge is still 
spinning - wait for the timer to stop and the audible alarm to 
sound.

5

After the unit has stopped, carefully remove the capillary tubes. 

NB - Wait until the rotor has stopped completely before 
opening the lid. 

Reading your Haematocrit results:

1

Place the reader card directly behind the tube and align the top 
and bottom of the fluid (not the tube or sealant) with the 0 and 
100 lines (see fig. 5).

2

The separation line between the red cells and plasma should be 
read on the graph at the point of separation.

3

NB - The separation line will be at a slight angle, therefore you 
must read the midpoint of the angle for an accurate percentage.

4

0 - 100 on the reader indicates a percentage (%) result. To express 
the result in SI units multiply the % value by 0.01, eg 45% = 0.45

5

SAFETY - Used capillary tubes must be disposed of as clinical 
waste in a sharps container.

Care & Maintenance

With proper care and maintenance the clinispin HCT centrifuge will 
provide years of service.

A)

Clean the inside of the centrifuge. Whenever there is a tube 
breakage or spillage, it is recommended that the inside of the 
centrifuge is cleaned, and if possible disinfected. 
(Do not immerse the electrical components in any liquid).

B)

The Clinispin HCT utilises a maintenance-free 
brushless motor and its bearings are permanently 
lubricated; it should not need servicing for the life of 
the centrifuge. Likewise, the electrical components 
were designed for high reliability and should not 
need regular service.