To edit the existed test items:
1) Click “Edit”.
2) The test items are listed on the left of the interface. Click on the test item to be edited, its name will 
    appear on the ProName box.
3) Click on the Min box, delete the original value and type in the newly set value;
4) Assign a new value to Max by following the same steps as 3);
5) Click “Save” to save the change, otherwise click “Quit”.

To restore original settings:
1) Click “Reset”.
2) Click “OK” to confirm restoring original settings, otherwise choose “Cancel”.

Device Parameters:
 To read the information about the following device parameters, click         ; 

AD magnification times: Magnificating times of fluorescence signal. 
Total length of the slot in running: The maximum distance running forward when motor starts with 
photoelectric switch.
Barcode recognition deadline: The timeout period of barcode recognition.
 Sampling starting point: The starting position of fluorescence data collection.

1. Run a Test
Run a Test to analysis a test device.
To start running a test: 
1) Insert ID chip into the ID chip port until resistance is felt. 
    It is recommended to insert the ID chip after successful booting.
    This step is necessary when a new lot of test device of specific test is applied on the meter for the first 
    time. Once the meter recognizes the Lot information of the chip, there is no need to read ID chip again 
    before running test, unless a new lot of test device is applied. 
    The ID chip can either be pulled out once the lot information is recognized or remain inserted.
2) Navigate to the main interface.
3) Click “Read ID Chip”. A message said “Warning: The ID chip ready” will prompt, click “Yes”. It 
    indicates that the ID chip can be recognized by the meter.
4) Choose test mode by click on “Quick Test” or “Standard Test”.
5) Choose Test mode.
    It is assumed that before switching Test Mode, the selector switch for operating mode, which is on the 
    left side of analyzer, is on MAC. 
    

Finecare™ FIA Meter Plus

 provides two test modes, the Standard Test mode and Quick Test mode. 

    Standard Test:  Push the test device into the test device tray right after the specimen is fully absorbed.  
    The analyzer acts as an incubation place for reaction for a preset time duration. Wait until the analyzer  
    counts down to zero, the test result will be shown on the screen. 
    Quick Test: Push the test device into the test device tray after the reaction completes.
6) Input detailed patient information by clicking “Detailed information”..
    If the user does not input a serial number for a record, the analyzer will automatically assign a serial 
     number for it.
7) Apply the test sample on the test device according to instructions of specific test. 
8) Start the test. 
    Quick Test: Gently push the test device that is completely reacted into the test device Tray. Click       .  
    Click “StartTest”.
    Standard Test: Gently push the test device into the test device tray immediately after the specimen is fully 
    absorbed. Click        . Click “StartTest”.
9) The meter will analyze the test device whilst the blue bar runs.

Page 

13

Page 

12

Section VI Operation

*Note:
These information are readonly to users. If you choose 
“Write”, a message said “Warning: Please enter 
password” will pop out, as users are not authorized to 
edit these parameters. 

Содержание Finecare FIA Meter Plus FS-113

Страница 1: ...Version 30 06 2016 Finecare FIA Meter Plus Operation Manual Model No FS 113...

Страница 2: ...Page 3 4 7 8 9 13 17 24 25 26 27 Content Introduction Contents Specifications Warning Precautions and Limitations Installation Operation Communication Settings Quality Control Maintenance and Care Tro...

Страница 3: ...more energy the fluorescent dye gives off the stronger the signal After a buffer mixed sample is added to the test device the test device is inserted into the Finecare FIA Meter Plus The meter measure...

Страница 4: ...Section Contents 1 List of Contents No 1 2 3 4 5 6 7 8 9 Quantity 1 1 1 1 1 1 1 1 1 Item Finecare FIA Meter Plus Power adapter including the power cord Quality Control Card Ethernet cable Printer pape...

Страница 5: ...ity Size Weight Miscellaneous Temperature Humidity Atmospheric pressure Location Temperature Humidity 10 50 14 122 85 10 30 50 86 10 80 86kPa 106kPa Dry clean flat horizontal surface away from direct...

Страница 6: ...ou Wondfo Biotech Co Ltd or its representative Do not place objects on Finecare FIA Meter Plus This may damage the optical alignment in turn result in compromised performance or mechanical damage The...

Страница 7: ...box 4 2 Set Up Temperature The symbol and the temperature shown below is an indicator of the temperature inside the meter 1 Click set up on the main interface 2 Input q1 in the Authorization codes fi...

Страница 8: ...r remain inserted 2 Navigate to the main interface 3 Click Read ID Chip A message said Warning The ID chip ready will prompt click Yes It indicates that the ID chip can be recognized by the meter 4 Ch...

Страница 9: ...r the tested item the test result the sample type Name of the patient and the time when the test is conducted are shown for each record If records are more than the screen can show slide up or down on...

Страница 10: ...LIS or a PC immediatedly after the test completes 2 history data stored in the meter to be sent to the LIS or a PC Transfer Protocol Instruction Finecare FIA Meter Plus sends data to outside by UDP cl...

Страница 11: ...access the Internet one end of the network cable connects Finecare FIA Meter Plus s Ethernet port the other end connects PC s Ethernet port of new Installed NIC Alter the new NIC IP address in compute...

Страница 12: ...ollowing information in the TCP IP address window Click Add See Fig 6 9 IP address 192 168 0 88 Subnet mask 255 255 255 0 4 Return Advanced TCP IP settings add gateway by clicking Add as indicated in...

Страница 13: ...he computer terminal See Fig 6 17 3 Open LIS data receiving software installed in the computer select network port Click Start See Fig 6 18 Note Choose UDP as communication networks Don t choose Seria...

Страница 14: ...been used in clinical setting and needs to be repackaged replaced or transported it must be cleaned Use disinfectant e g bleach concentration lower than 0 1 and a cloth to wipe external surface of th...

Страница 15: ...r ID chip Computational Abnormality Mechanical Failure Buzzing Noise Software Abnormality Electrostatic Influence Electric Circuit Failure Test device or ID chip was not inserted The Lot Number on the...

Страница 16: ......

Отзывы: