11.2
Air Temperature Readings
Error Code
Problem
Reason/Solution
E02
The air temperature value is
under the lower limit.
Put the meter in regular room temperature
for 30 minutes, if the error message still
appears, send the device for repair.
E03
The air temperature value
exceeds the upper limit.
Put the meter in regular room temperature
for 30 minutes, if the error message still
appears, send the device for repair.
E31
The temperature sensor or
measuring circuit is dam-
aged.
Send the device for repair.
11.3
Air Humidity Reading
Error Code
Problem
Reason/Solution
E04
The air temperature meas-
urement has an error code.
Refer to the above mentioned temperature
error code.
E11
The RH calibration has
failed.
Send the meter back for recalibration.
E34
The RH sensor or the
measuring circuit is dam-
aged.
Send the meter back for repair.
11.4
General
Error Code
Problem
Reason/Solution
E36
- Supply voltage too low
- Signal during Power
breakdown
-
Controll contact to the power supply