Wöhler FA 320

 

3.4 Air 

volume 

3.4.1 Setting 

the 

area 

Press and hold the MODE key for more than two seconds to enter the air volume 
mode. 
To measure the volume, the outlet size must be input first. 
There are three different ways to enter the size: 
 
Length x Width: 
This is the default choice. 
Press the HOLD key to select the setting digital first, then press the MX/MN key to 
select the appropriate value. 
After having entered the length, shortly press the ENTER key to enter into the 
width setting mode. Press ENTER again to finish the size setting and enter the 
measuring mode (see 3.4.2). 
 
Diameter: 
While in default outlet setting mode (length), press the ENTER key for over two 
seconds to choose the diameter setting. 
Press the HOLD key to select the setting digital first, then press the MX/MN key to 
select the appropriate value. 
After this selection, shortly press ENTER again to finish the size setting and enter 
the measuring mode (see 3.4.2). 
 
Area: 
While in the diameter setting mode, press the ENTER key for over two seconds to 
enter the area setting. 
Press the HOLD key to select the setting digital first, then press the Mx/Mn key to 
select the appropriate value. 
After this selection, shortly press ENTER again to finish the size setting and enter 
the measuring mode (see 3.4.2). 

Содержание 22107

Страница 1: ...er Art Nr 22107 Contents 1 Preface 12 2 Specifications 13 3 Operations 14 4 Declaration of Confirmity 19 5 Guarantee and Service 20 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1...

Страница 2: ...fore com missioning the unit In principle the W hler FA 320 Fan Anemometer should only be used by skilled personnel for its intended purpose and within the specified range of data Any liability or war...

Страница 3: ...age determination of eight measuring points it will be possible to measure even large outlets In this case the air volume flow will be calculated automatically The user can select between different un...

Страница 4: ...al Data Dimensions vane 170 x 77 x 40 mm Dimensions meter 175 x 70 x 33 mm Weight vane 150 g Weight meter 185 g Power 4 x 1 5 V battery AAA 3 Operation 3 1 Important Reminder Rotate the cover before m...

Страница 5: ...oint wet bulb relative humidity velocity 3 2 3 ENTER Press ENTER to confirm the setting and calibration 3 2 4 Mx Mn UP Press to view MAX MIN value Press to select the value of each digital cycle 3 2 5...

Страница 6: ...press the POWER key for more than two seconds to enter the unit selection mode Press or to select the unit and then press ENTER to save C RH m s m m cm or F RH ft m cfm inch 3 3 4 Velocity with averag...

Страница 7: ...setting and enter the measuring mode see 3 4 2 Diameter While in default outlet setting mode length press the ENTER key for over two seconds to choose the diameter setting Press the HOLD key to selec...

Страница 8: ...eps when the 60 seconds are over After the time is up the 60 seconds average volume will be indicated in the dis play If there is more than one outlet to be measured press the ENTER key again to repea...

Страница 9: ...ntial protection requirements of Council directive 89 336 EC 92 31 EEC and 93 68 EEC on the approximation of the laws of the Member States relating to electromagnetic compability The meter conforms to...

Страница 10: ...centers listed below You can obtain immediate help from our engineers by telephone 5 3 Information on disposal You may hand in any defective batteries taken out of the unit to our company as well as...

Отзывы: