background image

12

Technical Data:

Technology

Microprocessor

Programs

Button:

Start-Stop with timer, 60 min

Foot pedal:

Standard (flame during pressed foot pedal
or activated external IR motion sensor, foot
pedal and external IR motion sensor optional)

Safety features

Safety Control System (SCS)
with gas safety cut off:

ignition  and  flame  control
temperature monitor,
burner head clogging monitor (BHC)
automatic unit switch off, 4 h

Gas supply and consumption

Gas connection:

1/4” left with gas filter

Gas types:

II

2ELL3B/P

: natural gas E/LL,18 - 25 mbar

liquid gas, 20 - 50 mbar

Connected load:

43 g/h liquid gas

53 l/h natural gas (E)
49 l/h natural gas (LL)

Continuous cartridge operation:

CV 360 - 65 min,

CG 1750 - 210 min,

(approx.)

C 206 - 230 min,

CP 250 - 305 min,

CV 470 - 550 min,

CV 270 - 280 min

Temperatures

Flame temperature:

1200°C on liquid gas
1170 °C on natural gas (E)

Temperature threshold level:

0.55 kW liquid gas, 0.50 kW natural gas (E),
0.40 kW natural gas (LL)

Electrical

Power consumption:

2 VA

Power connection:

100 - 240 V / 50/60 Hz / max. 0.3 A
9 V DC / 1 A

Mechanical

Burner tube:

removable, stainless steel

Measurements (w x h x d):

89 x 34 x 88 mm

Height with burner tube:

94 mm

Weight:

385 g

Licenses

DIN-DVGW Reg.-No.:

NG-2211AS0167

CE:

EN 61326-1, EN 61010

EC guidelines:

2004/108/EC, 2006/95/EC

Содержание Flame 100

Страница 1: ...1 EN Sets new standards in the Bunsen burner class Instruction manual DIN30665 1...

Страница 2: ...foot pedal 7 Power connector for 9 V DC 8 Gas inlet R 1 4 L gas adapter lefthandthread 9 Holding device for inoculationloopholder 10 Electrodeholder 10a Screw for Electrode holder 11 Monitorelectrode...

Страница 3: ...2 4 Foot pedal connection External IR motion sensor 7 3 Operation 7 3 1 On Off switch 7 3 2 Application programs 7 3 3 Flame regulation 7 3 4 Switch off End of work 7 4 Safety Symbols Safety Control S...

Страница 4: ...rs Check the conditionofthetube hosefrequently Dependingupontypeoftube hose hose clamps are required All gas connections must be adequately tightened with two wrenches Ensure gasproofnesswithasuitable...

Страница 5: ...rner tube 5 by turning counter clockwise and unscrew it completely Loosen the nozzle N 55 14 for natural gas with the wrench L1 supplied by turning counter clockwise unscrew completely and remove Ther...

Страница 6: ...or other gas approved pressure regulator 50mbar must be used for liquid gas Pay attention to your relevant rules for using liquid gas Several gas cartridge adapter are optionally available 7 8 6 2 3...

Страница 7: ...Foot pedal connection External IR motion sensor Insert the connection cable of the foot pedal or the external IR motion sensor accessory into the socket 6 at the back of the unit The foot pedal extern...

Страница 8: ...off and the unit indicates a malfunction see paragraph 5 5 5 Error displays Ignition failure Status LED 4 blinks red 2x This signal appears and indicates a malfunction if the flame fails to ignite af...

Страница 9: ...r wash it in a dishwasher To remove the burner tube proceed as follows Loosen the screw for the electrode holder 10a with the flat head screwdriver supplied L3 Slightly tilt the electrode holder 10 ba...

Страница 10: ...against any manufacture defects in material and workmanship The WLD TEC warranty guarantees all gas burners under normal usage conditions and does not cover any damages as a direct result of user misu...

Страница 11: ...ollowing standards togetherwiththenormativedocumentA2mainsconnection 1 ElectromagneticCompatibilityDirective 1 1 EN 61326 1 2006 Electricalequipmentformeasurement controland laboratoryuse EMCrequireme...

Страница 12: ...bar liquid gas 20 50 mbar Connected load 43 g h liquid gas 53 l h natural gas E 49 l h natural gas LL Continuous cartridge operation CV 360 65 min CG 1750 210 min approx C 206 230 min CP 250 305 min C...

Страница 13: ...tube 5 Especially remove substances in the marked areas at the electrodes If there are contaminants in this area the flame cannot encircle the electrodes correctly Clean this areas with a brush The b...

Страница 14: ...ee paragraph 5 Status LED blinks red 4x Overtemperature see paragraph 5 Status LED flashes amber BHC activ see paragraph 4 Service address WLD TEC GmbH Production Service Halle Kasseler Str 49 37318 A...

Страница 15: ...G ttingen Telefon 49 0 551 793789 Fax 49 0 551 793707 Internet http www wld tec com ProductionundService Halle KasselerStra e49 D 37318Arenshausen Telefon 49 0 36081 68940 Fax 49 0 36081 68942 Email...

Отзывы: