background image

 

 

Safety Lock 

Dispensing position 

4 Functional characteristics 

4.1 

Air purging

 

Before starting to pump, it is essential to place a collecting 
vessel under the cannula. The valve axle has to be in the 
dispensing position 

(arrow up)

.  

Now perform short pumping movements until no air 
bubbles are leaving the cannula. Perform regular and 
smooth movements and make sure that the outer housing 
(2) is always pushed all the way down.   

 
 

4.2 Volume adjustment 

The quick-volume-adjustment is performed with the volume rocker switch (3).  
To set the required dispensing quantity, hold the rocker switch pressed and 
move the arrow to the desired volume. Now unhand the rocker and make sure it 
fits correctly into the sprockets. (For 

Labmax 100 ml

 and 

250 ml

: turn 

adjustment screw counterclockwise until the arrow is easy to move, choose 
your desired volume and then fix the adjustment screw.) 
 

4.3 Dispensing 

Avoid personal injury from chemicals. To protect your self wear eye protection 
and use appropriate safety equipment and clothing. 

Please follow all safety instructions and observe operating procedures. 
Raise the outer housing (2) until it stops at the point set by the volume rocker 
switch. The set quantity will be dispensed into the collecting vessel by pressing 
down the outer housing (2) to the lowest point. The movements should be 
smooth and constant to achieve an exact ejection volume. 
 

4.4 Safety-Lock 

 
 
Turn the cannula to the left stop 

(arrow down)

.  

The liquid into the cannula will now rinse back into the bottle. 
Any further dispensing of liquid is not possible. 

 

 

5

5 Cleaning and maintenance 

5.1 Description of the components 
 

Volume rocker 
switch (3) 

Fix screw (1) 

Adaptor ring (4) 

Outer housing (2) 

Cap nut (5) 

Cannula holder (7) 

Ejection cannula (6) 

Valve head (19) 

Return tube (21) 

Suction tube (20) 

Adaptor ring (4) 

Cylinder (18) 

Cap nut (5) 

Cylinder Set (16) 

Plunger (17) 

Ejection valve (11) 

Suction valve (8) 

Valve star (9) 

Valve ball (10) 

edges on top! 

Содержание LABMAX OE

Страница 1: ...GmbH Am Bildacker 16 D 97866 Wertheim Main Phone 49 0 9342 9301 0 Fax 49 0 9342 9301 77 Email info witeg de www witeg de Version 06042009 Operating Manual for Dispenser max OE OE S OE HF Please study...

Страница 2: ...ling chemicals e g protective clothing protective goggles Use the Labmax only regarding to the resistance of materials and for the purpose for which is intended Always check the device for leaks fixed...

Страница 3: ...f substances catalysed by platinum iridium alloys The temperature of the Labmax OE and reagent should not exceed 40 C 104 F 3 21 20 2 General product description The Labmax OE is witeg s economy dispe...

Страница 4: ...ustment screw 4 3 Dispensing Avoid personal injury from chemicals To protect your self wear eye protection and use appropriate safety equipment and clothing Please follow all safety instructions and o...

Страница 5: ...assemble the Labmax until you can surely clean it from all reagents Sometimes it could last out to rinse the device The cleaning can be performed with distilled water or alcohol The Labmax should be r...

Страница 6: ...the closing cap 13 This point is not applicable for Labmax 250 ml 3 Place the valve ball 10 and the valve star 9 into the suction valve Check that the spikes of the valve star 9 point upwards For Labm...

Страница 7: ...cap 13 2 5 100 ml 5 370 380 11 O Ring a 14 a 2 5 100 ml 5 370 381 O Ring b 14 b 2 5 10 ml 5 370 387 Cylinder set 16 2 5 ml 5 370 255 OE includes 4 5 18 5 ml 5 370 256 OE 10 ml 5 370 257 OE 25 ml 5 370...

Страница 8: ...ass tube with plastic coating for calcium chloride ensures dry atmosphere in storage vessel Article No 5 377 310 3 Filter against dust and bacteria in storage vessel pore size 2 m 4 5 m Article No 8 2...

Отзывы: