Stacking packages that weigh more than 60 kg/m² is completely prohibited, regardless of position. 

 
Dear Sir, Madam, 
 
Thank you for choosing a WINNCARE medical accessory. 
 
Our beds and accessories are designed and manufactured in compliance with the applicable essential 
requirements, according to European Directive 2007/47/EC. 
 
The commercial configurations, including boards and accessories manufactured by us, are tested in 
compliance with IEC 60601-2-52 (2009)

 

to ensure maximum safety and performance. 

 
Consequently, adhering to WINNCARE’s recommended usage conditions, and using original boards and 
accessories, is a condition of upholding the clauses of the guarantee and ensures that the fixed footboard 
bed rail is used safely. 
 
 

1. TRANSPORT CONDITIONS 

 
During transport, rails should be strapped and protected by packaging. 
 

NB

: stacking packages on the rail is completely prohibited. 

 
 
 
 

2. STORAGE CONDITIONS 

 
The  rails  must  be  stored  at  an  ambient  temperature  between  -10  °C  and  +50  °C,  with  relative  humidity 
between 30% and 75%. 
 
Atmospheric pressure between 700 hPa and 1060 hPa

 

, adhering to the same conditions as during transport. 

 
 
 
 
 
 
 
 
 

 

 

Adhere to the specified 

environmental conditions 

Содержание A671-00

Страница 1: ...h 4 Le Pas du Château 85670 SAINT PAUL MONT PENIT TEL 33 0 2 51 98 55 64 FAX 33 0 2 51 98 59 07 Email info medicatlantic fr Website http www winncare fr USER MANUAL fixed footboard bed rail A671 00 RIGHT RAIL A672 00 LEFT RAIL ...

Страница 2: ...0 FIXED RAILS 5 4 USAGE CONDITIONS 7 4 1 FUNCTION 7 4 2 PRECAUTIONS 7 4 3 RESIDUAL RISKS 8 5 MAINTENANCE CONDITIONS 8 5 1 IDENTIFICATION 8 5 2 MAINTENANCE 8 5 3 CLEANING 8 5 4 DISINFECTION 8 5 4 1 OBJECTIVE 8 5 4 2 INDICATION 8 5 4 3 EQUIPMENT 8 5 4 4 TECHNIQUE 9 5 5 SHELF LIFE 9 5 6 GUARANTEES 10 6 SCRAPPAGE CONDITIONS 10 ...

Страница 3: ... and performance Consequently adhering to WINNCARE s recommended usage conditions and using original boards and accessories is a condition of upholding the clauses of the guarantee and ensures that the fixed footboard bed rail is used safely 1 TRANSPORT CONDITIONS During transport rails should be strapped and protected by packaging NB stacking packages on the rail is completely prohibited 2 STORAG...

Страница 4: ...671 00 A672 00 AL A671 00 A672 00 MS ML A671 00 A672 00 AE A671 00 A672 00 AF A671 00 A672 00 AG A671 00 A672 00 AH A671 00 A672 00 AX A671 00 A672 00 XA A671 00 A672 00 XB A671 00 A672 00 Beds compatible with the FIXED RAIL Fixed rail Foot of the bed Head of the bed Folding rail The fixed rail is not compatible with DUO 140 160 beds or the PITCHOUNE KALIN bed The fixed rail is not compatible with...

Страница 5: ... 3 1 Assembling and installing the A671 00 A672 00 fixed rails Using a 13 mm socket wrench remove the board mount and replace with the new part Attach the footboard to the bed Bed rail lock mechanism 1 2 ...

Страница 6: ...6 Attach the fixed rail to the bed Attach the screw and the plastic nut to lock the rail in place 3 4 ...

Страница 7: ...head neck or chest When deciding whether or not to use the fixed footboard bed rails a risk benefit ratio assessment should be first be performed This assessment should take account of the department s monitoring ability A standardized institutional protocol for using the fixed footboard bed rail can be prepared the patient s physical and mental health their needs abilities lucidity size and level...

Страница 8: ... avoid transmitting bacteria from one patient to another Remove any organic substance via Mechanical action scrubbing Chemical action disinfection 5 4 2 Indication Rails are physically and bacteriologically clean 5 4 3 Equipment Microfibre wipes Detergent or Detergent Disinfectant CE marked Surface DD and Surface Disinfectant Chlorine derivatives 30 alcohol NB undiluted DD products and bleach shou...

Страница 9: ...actuators electrical components and remote controls with microfibre wipes soaked with disinfectant Do not rinse or dry Check for normal function and the various operations Repair if necessary NB Take the measures recommended by the hospital s internal hygiene department in case of additional precautionary measures Contact Droplets Air precautions Using a greater than 5000 ppm bleach solution 0 5 a...

Страница 10: ... label Replacement is made by providing original parts within the limit of the guarantee period with our resale network determining the start of the guarantee period Returning defective parts is mandatory so that the guarantee can be applied correctly and to avoid receiving an invoice 6 SCRAPPAGE CONDITIONS The product must be scrapped if it no longer meets the essential requirements particularly ...