Positioning the side rails A579-00 / A580-00 / A581-00
poorly may undermine safety and cause a malfunction.
Side rails
must not be used when patients are not adults or are
146 cm tall.
The user and staff must be trained and aware of the risks associated with using the bed. He
must not allow it by children and be vigilant when used by confused or disoriented people.
There must be at least 220 mm between the top of the side rail and uncompressed and no
therapeutic mattress surface. It will be advisable to tend towards this specification in the case
of the use of a therapeutic mattress.
Incorrect positioning of the A680-00 barriers
can lead to safety hazards or malfunctions.
These
barriers are used for children with a height of between 125 cm and 145 cm for the
“KALIN” junior medical bed.
4.4. Precautions for use
The side rails must be used by someone outside of the bed, who must ensure that nothing or nobody is situated
in the area in which the bed is being used.
Side rails are a medical device, and in this regard we inform you that it must not be modified under any
circumstances. You must ensure its traceability. If you assemble different types of medical devices, you are
responsible for conducting the risk analysis and making the CE declaration.
We recommend putting the bed in its low position after every use and while the patient is resting, to reduce the
height of falls by a confused or agitated person.
For assistance, if necessary, in mounting, operation or maintenance or to report unexpected operation or
events, call your supplier or Winncare.
4.5. Residual risks
DANGER
RESULT
REQUIREMENTS FOR THE USER
Trapping
Hands pinched
Handle barrier by the upper bar
Trapping
Crushing body or object
Check before lowering the barrier that
the movement area is free
5. MAINTENANCE CONDITIONS
5.1. Identification
5.2. Upkeep
Manufacturer's address and
contact details
Warning
Complies with regulations (EU)
Product reference and name
Read the instruction manual
Product serial number
YY : Year of manufacture
MM : Month of manufacture
XXXXX :Internal numbering
PP : Product empty weight
Color code and designation
Medical device)