Document No.: TMM-2220-13
Revision: A
Page 8 of 30
NOTICES:
MAXIMUM PATIENT WEIGHT IS 500 lbs (227 kg)
•
Maximum patient weight capacity of chair is 500 pounds (inclusive of clothing, accessories, etc.). If
exceeded, damage to chair could occur.
CHAIR IS INTENDED FOR INDOOR USE ONLY
•
To ensure proper operation and extend chair life, only use chair in specified environment.
USE CDC’s
UNIVERSAL PRECAUTIONS
•
When maintaining chair after clinical use, service personnel must use UNIVERSAL PRECAUTIONS as
defined by CDC (Centers for Disease Control and Prevention).
MAINTAIN CHAIR REGULARLY
•
To ensure proper operation and extend chair life, inspect, maintain, and service chair on a regular basis.
Inspection, maintenance, and service details are located later in this manual.
REMOVE BATTERY DURING CHAIR STORAGE
•
For “B” Option Only
If chair is to be stored or not in use for four (4) days or more, remove battery pack from chair. Refer to
Battery Charger Installation and Usage Instructions (IM TMA57-15), under the “Cautionary Usage
Instructions” section, for more information.
•
For “A” Option Only
If chair is to be stored or not in use for four (4) days or more, leave chair plugged into an AC wall outlet or
unplug battery from control box. For more information, refer to “A” Option section below.
SURGICAL SUPPORT LEG MUST BE “UNLOCKED” TO ACTUATE BACK AND TILT MOTORS
•
To prevent chair damage, the back and tilt motors are disengaged while the surgical support leg is in the
locked position. If you need to adjust the back/seat angle return the surgical support leg lever to the
“unlocked” position, adjust the back/seat, and then re-lock the surgical support leg lever.