WILDLIFE 265A Скачать руководство пользователя страница 4

IOP 006A.1 

Effective Date: 1/09/2017 

Page 4 of 4 

 

unit. 

2. 

Launch SW109 and turn the logger on. 

 

3. 

SW109 and logger are in communication when the “LOG” icon on the SW109 

menu is yellow.  If not, click “COM Port” on the Configuration pull down menu 

and make sure the appropriate COM port is selected (usually COM1) and the 

baud rate matches the baud rate set on the logger (4800). 

 

4. 

Click the yellow “LOG” icon to download the logger data and save the file 

(filename.ply).  

 

 

VII. 

Orion 9109WL Triode storage and maintenance 

 

A. 

Short-term storage (up to one week):  Cover electrode with the protective cap containing 

a small wad of cotton saturated in ph electrode storage solution (Orion 910001) or pH 4 

buffer. 

 

B. 

Long-term storage (more than one week):  Rinse off any salt deposits with distilled water.  

Cover with protective cap containing a few drops of electrode storage solution (or ph 4 

buffer). 

 

C. 

Before using after long-term storage, rinse off any salt deposits with distilled water.  

Immerse in electrode storage solution for at least one hour. 

 

D. 

The electrode can be cleaned by soaking in pH 4 buffer.  If this does not improve 

performance, use the ph electrode cleaning kit (Orion 900020), following instructions in 

the kit.      

 

MAINTENANCE:

 

 

 

All sensors are shipped with the measuring end covered with a wetting cap.  Sensors are stored in a pH 4 

 

buffer solution when not in use.  Vertical storage is recommended to prevent leakage of the buffer solution 

 

from the wetting cap. 

 

A number of other maintenance and troubleshooting procedures are described in the instrument and sensor 

instruction manuals:  Models 265A and 266S Waterproof and Intrinsically Safe Portable pH Meters 

 

 

REFERENCE 

 

Instruction manual for the Models 265A and 266S Waterproof and Intrinsically Safe Portable pH Meters 

 

 

 

This procedure has been reviewed and approved by the undersigned representative of the U.S. 

Fish and Wildlife Service 

 

 

REVIEWED/APPROVED______________________________     DATE____________ 

 

 

 

 

      Field Supervisor (U.S.) 

 

Jessica M. Barber 

Digitally signed by JESSICA 

BARBER 

Date: 2020.03.10 10:40:12 

-04'00'

Содержание 265A

Страница 1: ...Oceans Sea Lamprey Control Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada and U S Fish and Wildlife Service Lake Champlain Fish and Wildlife Resources Office 11 Lincoln St Essex Junction VT 05452 INSTRUMENT OPERATING PROCEDURE INSTRUMENT Portable pH measurement and logging device MODEL 265A 266S MANUFACTURER Thermo Orion PRECAUTIONS POTENTIAL INTERFERENCES Assure that the pro...

Страница 2: ...se probe in pH 7 buffer stir buffer for several seconds to remove bubbles from electrode and wait a minimum of 5 minutes for the probe to stabilize Press cal and gently stir to calibrate to pH 7 the measured mV value is displayed When pH 7 calibration is complete CAL 2 is displayed G Rinse probe with distilled water pat off excess water and rinse with the pH 4 buffer Immerse probe in pH 4 buffer s...

Страница 3: ...gger memory and begins data logging The first logged data point is in memory location 00 D The Orion model 265A and 266S records up to 100 data points memory locations 00 to 99 The data log does not automatically stop after reaching 100 data points It continues logging beginning at location 00 and overwrites previously recorded data V Deployment A Place the unit in an inconspicuous location on the...

Страница 4: ...any salt deposits with distilled water Immerse in electrode storage solution for at least one hour D The electrode can be cleaned by soaking in pH 4 buffer If this does not improve performance use the ph electrode cleaning kit Orion 900020 following instructions in the kit MAINTENANCE All sensors are shipped with the measuring end covered with a wetting cap Sensors are stored in a pH 4 buffer solu...

Отзывы: