WILDLIFE 265A Скачать руководство пользователя страница 1

IOP 006A.1 

Effective Date: 1/09/2017 

Replaces IOP 006A.0 

Page 1 of 4 

 

 

 

 

 

 

 

 

U.S. Fish and Wildlife Service 

 

 

 

 

 

 

 

Marquette Biological Station 

 

 

 

 

 

 

 

3090 Wright Street 

 

 

 

 

 

 

 

Marquette, Michigan  49855 

 

 

 

 

 

 

 

U.S.A. 

 

 

 

 

 

 

 

 

 

and 

 

 

 

 

 

 

 

 

U.S. Fish and Wildlife Service 

 

 

 

 

 

 

 

Ludington Biological Station 

 

 

 

 

 

 

 

5050 Commerce Drive 

 

 

 

 

 

 

 

Ludington, Michigan  49431  

 

 

 

 

 

 

 

U.S.A. 

 

 

 

 

 

 

 

 

 

and 

 

 

 

 

 

 

 

 

Department of Fisheries and Oceans 

 

 

 

 

 

 

 

Sea Lamprey Control Centre 

 

 

 

 

 

 

 

1219 Queen Street East 

 

 

 

 

 

 

 

Sault Ste. Marie, Ontario P6A 2E5 

 

 

 

 

 

 

 

Canada 

 

 

 

 

 

 

 

 

 

and  

 

 

 

 

 

 

 

 

U. S. Fish and Wildlife Service 

Lake Champlain Fish and Wildlife Resources Office 

 

 

 

 

 

 

 

11 Lincoln St. 

Essex Junction, VT 05452 

 

 

INSTRUMENT OPERATING PROCEDURE

 

 

INSTRUMENT:

 

 

 

Portable pH measurement and logging device 

 

MODEL: 

 

 

265A/266S 

 

MANUFACTURER: 

 

 

Thermo Orion

 

 

PRECAUTIONS: 

 

 

POTENTIAL INTERFERENCES 

 

 

Assure that the probe is clean, free from scratches or organic build-up. 

 

Store the pH probe wet in probe storage solution for long periods. 

 

 

SAFETY

 

 

 

 

 

 

Содержание 265A

Страница 1: ...Oceans Sea Lamprey Control Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada and U S Fish and Wildlife Service Lake Champlain Fish and Wildlife Resources Office 11 Lincoln St Essex Junction VT 05452 INSTRUMENT OPERATING PROCEDURE INSTRUMENT Portable pH measurement and logging device MODEL 265A 266S MANUFACTURER Thermo Orion PRECAUTIONS POTENTIAL INTERFERENCES Assure that the pro...

Страница 2: ...se probe in pH 7 buffer stir buffer for several seconds to remove bubbles from electrode and wait a minimum of 5 minutes for the probe to stabilize Press cal and gently stir to calibrate to pH 7 the measured mV value is displayed When pH 7 calibration is complete CAL 2 is displayed G Rinse probe with distilled water pat off excess water and rinse with the pH 4 buffer Immerse probe in pH 4 buffer s...

Страница 3: ...gger memory and begins data logging The first logged data point is in memory location 00 D The Orion model 265A and 266S records up to 100 data points memory locations 00 to 99 The data log does not automatically stop after reaching 100 data points It continues logging beginning at location 00 and overwrites previously recorded data V Deployment A Place the unit in an inconspicuous location on the...

Страница 4: ...any salt deposits with distilled water Immerse in electrode storage solution for at least one hour D The electrode can be cleaned by soaking in pH 4 buffer If this does not improve performance use the ph electrode cleaning kit Orion 900020 following instructions in the kit MAINTENANCE All sensors are shipped with the measuring end covered with a wetting cap Sensors are stored in a pH 4 buffer solu...

Отзывы: