background image

WILDEN PUMP & ENGINEERING, LLC  

22 

WIL-10131-E-01 

Pump will not run or runs slowly.

1. 

Ensure that the air inlet pressure is at least 0.4 bar (5 psig) 
above startup pressure and that the differential pressure 
(the difference between air inlet and liquid discharge 
pressures) is not less than 0.7 bar (10 psig).

2.  Check air inlet filter for debris (see recommended instal-

lation).

3.  Check for extreme air leakage (blow by) which would 

indicate worn seals/bores in the air valve, pilot spool and 
main shaft.

4.  Disassemble pump and check for obstructions in the air 

passageways or objects which would obstruct the move-
ment of internal parts.

5.  Check for sticking ball check valves. If material being 

pumped is not compatible with pump elastomers, swell-
ing may occur. Replace ball check valves and seals with 
proper elastomers. Also, as the check valve balls wear 
out, they become smaller and can become stuck in the 
seats. In this case, replace balls and seats.

6.  Check for broken inner piston which will cause the air 

valve spool to be unable to shift.

7.  Remove plug from pilot spool exhaust.

Pump runs but little or no product flows.

1.  Check for pump cavitation; slow pump speed down to 

allow thick material to flow into the liquid chambers.

2.  Verify that vacuum required to lift liquid is not greater 

than the vapor pressure of the material being pumped 
(cavitation).

3.  Check for sticking ball check valves. If material being 

pumped is not compatible with pump elastomers, swell-
ing may occur. Replace ball check valves and seals with 
proper elastomers. Also, as the check valve balls wear 
out, they become smaller and can become stuck in the 
seats. In this case, replace balls and seats.

Pump air valve freezes.

1.  Check for excessive moisture in compressed air. Either 

install a dryer or hot air generator for compressed air. 
Alternatively, a coalescing filter may be used to remove 
the water from the compressed air in some applications.

Air bubbles in pump discharge.

1.  Check for ruptured diaphragm.
2.  Check tightness of outer pistons. (Refer to Section 8C.)
3.  Check tightness of clamp bands and integrity of o-rings 

and seals, especially at intake manifold.

4.  Ensure pipe connections are airtight.

Product comes out air exhaust.

1.  Check for diaphragm rupture.
2.  Check tightness of outer pistons to shaft.

OPERATION: The P8 and PX8 are pre-lubricated, and do 
not require in-line lubrication. Additional lubrication will not 
damage the pump, however if the pump is heavily lubricated 
by an external source, the pump’s internal lubrication may be 
washed away. If the pump is then moved to a non-lubricated 
location, it may need to be disassembled and re-lubricated as 
described in the ASSEMBLY/DISASSEMBLY INSTRUCTIONS.

Pump discharge rate can be controlled by limiting the volume 
and/or pressure of the air supply to the pump (preferred 
method). An air regulator is used to regulate air pressure. 
A needle valve is used to regulate volume. Pump discharge 
rate can also be controlled by throttling the pump discharge 
by partially closing a valve in the discharge line of the pump. 
This action increases friction loss which reduces flow rate. 
(See Section 5.) This is useful when the need exists to control 
the pump from a remote location. When the pump discharge 
pressure equals or exceeds the air supply pressure, the 
pump will stop; no bypass or pressure relief valve is needed, 
and pump damage will not occur. The pump has reached a 
“deadhead” situation and can be restarted by reducing the 
fluid discharge pressure or increasing the air inlet pressure. 

The Wilden P8 and PX8 pump run solely on compressed 
air and does not generate heat, therefore your process fluid 
temperature will not be affected.

MAINTENANCE AND INSPECTIONS:  Since each application 
is unique, maintenance schedules may be different for every 
pump. Frequency of use, line pressure, viscosity and abra-
siveness of process fluid all affect the parts life of a Wilden 
pump. Periodic inspections have been found to offer the best 
means for preventing unscheduled pump downtime. Person-
nel familiar with the pump’s construction and service should 
be informed of any abnormalities that are detected during 
operation.

RECORDS: When service is required, a record should be 
made of all necessary repairs and replacements. Over a 
period of time, such records can become a valuable tool for 
predicting and preventing future maintenance problems and 
unscheduled downtime. In addition, accurate records make 
it possible to identify pumps that are poorly suited to their 
applications.

S U G G E S T E D   O P E R A T I O N   &   M A I N T E N A N C E

T R O U B L E S H O O T I N G

Содержание Original P8

Страница 1: ...S i m p l i f y y o u r p r o c e s s Engin e e ri n g Operati o n Mainte n a n ce Original Series PLASTIC Pumps P8 PX8 WIL 10131 E 01...

Страница 2: ...TIC Performance Operating Principal 10 How to Use this EMS Curve 11 Performance Curves Rubber Fitted 14 TPE Fitted 15 PTFE Fitted 16 Ultra Flex Fitted 17 Suction Lift Curve 18 SECTION 6 SUGGESTED INST...

Страница 3: ...es and containers must be grounded when handling flammable fluids and whenever discharge of static electricity is a hazard To ground the Wilden Champ all clamp bands must be grounded to a proper groun...

Страница 4: ...MS BNS BUNA N Red Dot BNU BUNA N ULTRA FLEX EPS EPDM Blue Dot EPU EPDM ULTRA FLEX FSS SANIFLEX Hytrel Cream NES NEOPRENE Green Dot NEU NEOPRENE ULTRA FLEX PUS POLYURETHANE Clear TEU PTFE W EPDM BACK U...

Страница 5: ...is on its suction stroke air behind the diaphragm has been forced out to the atmosphere through the exhaust port of the pump The movement of diaphragm B toward the center block of the pump creates a v...

Страница 6: ...6 METRIC mm STANDARD inch P 122 DIA 4 8 DIA R 152 DIA 6 0 DIA S 20 DIA 0 8 DIA S e c t i o n 4 D I M E N S I O N A L D R A W I N G S P8 Plastic PX8 Plastic DIMENSIONS ITEM METRIC mm STANDARD inch A 4...

Страница 7: ...ly operation parameters will fall in the center of the pump performance curve S e c t i o n 5 A P E R F O R M A N C E P8 PLASTIC RUBBER FITTED Height 770 mm 30 3 Width 490 mm 19 3 Depth 333 mm 13 1 Es...

Страница 8: ...ed so that daily operation parameters will fall in the center of the pump performance curve Water Discharge Flow Rates LPM P8 PLASTIC ULTRA FLEXTM FITTED Height 770 mm 30 3 Width 490 mm 19 3 Depth 333...

Страница 9: ...WIL 10131 E 01 7 WILDEN PUMP ENGINEERING LLC PTFE Diaphragms P8 PLASTIC SUCTION LIFT CAPABILITY S e c t i o n 5 B S U C T I O N L I F T C U R V E S...

Страница 10: ...WILDEN PUMP ENGINEERING LLC 8 WIL 10131 E 01 N O T E S...

Страница 11: ...PX8 P L A S T I C P X 8 P L A S T I C P E R F O R M A N C E...

Страница 12: ...igher performance lower operational costs and flexibility that exceeds previous industry standards Pro Flo XTM Operating Principal S e c t i o n 5 B Turning the dial changes the relationship between a...

Страница 13: ...in this case 9 8 scfm Step 2 Determining flow and air X Factors Locate your discharge pressure 40 psig on the verti cal axis of the EMS curve Figure 2 Follow along the 2 8 bar 40 psig horizontal line...

Страница 14: ...0 psig discharge pressure After locat ing this point on the flow curve draw a verti cal line downward until reaching the bottom scale on the chart and identify the flow rate In our example it is 38 6...

Страница 15: ...is different than the flow point plotted in example 2 1 After estimating or interpolating this point on the curve draw a vertical line downward until reaching the bottom scale on the chart and identif...

Страница 16: ...ed by 31 while the air consumption was reduced by 47 thus providing increased effi ciency For a detailed example for how to set your EMS see beginning of performance curve section Caution Do not excee...

Страница 17: ...by 10 while the air consumption was reduced by 19 thus providing increased effi ciency For a detailed example for how to set your EMS see beginning of performance curve section Caution Do not exceed 8...

Страница 18: ...d by 65 while the air consumption was reduced by 83 thus providing increased effi ciency For a detailed example for how to set your EMS see beginning of performance curve section Caution Do not exceed...

Страница 19: ...reduced by 40 while the air consumption was reduced by 69 thus providing increased effi ciency For a detailed example for how to set your EMS see beginning of performance curve section Caution Do not...

Страница 20: ...WILDEN PUMP ENGINEERING LLC 18 WIL 10131 E 01 PSIG BAR Dry Vacuum Inlet Air Pressure METER FT H20 PX8 PLASTIC SUCTION LIFT CAPABILITY S e c t i o n 5 D S U C T I O N L I F T C U R V E...

Страница 21: ...WIL 10131 E 01 19 WILDEN PUMP ENGINEERING LLC S e c t i o n 5 C N O T E S...

Страница 22: ...g pump performance Pumping volume can be set by count ing the number of strokes per minute and multiplying by displacement per stroke Sound levels are reduced below OSHA specifications using the stand...

Страница 23: ...oning valve will stop the air supply to the pump therefore stopping output This shut off valve should be located far enough away from the pumping equipment such that it can be reached safely in an eme...

Страница 24: ...r exhaust 1 Check for diaphragm rupture 2 Check tightness of outer pistons to shaft OPERATION The P8 and PX8 are pre lubricated and do not require in line lubrication Additional lubrication will not d...

Страница 25: ...fects of contact with your process fluid The Wilden P8 and PX8 have a 51 mm 2 inlet and outlet and are designed for flows up to 587 lpm 155 gpm Its air distribution system is based on a revolutionary...

Страница 26: ...the intake manifold to the liquid chambers Step 6 Figure 6 Lift intake manifold from liquid chambers and center section to expose intake valve balls and seats Inspect ball cage area of liquid chamber...

Страница 27: ...ire assembly can be removed from the center section Figure 12 2 The outer piston diaphragm and inner piston separate from the shaft which remains connected to the opposite side diaphragm assembly Figu...

Страница 28: ...P8 to run lube free Constructed of Acetal the Pro Flo air distribution system is designed to perform in on off non freezing non stalling tough duty applications TOOLS REQUIRED 3 16 Hex Head Wrench 7 3...

Страница 29: ...air chamber bolts with 7 32 hex head wrench Step 8 Figure 8 Remove pilot spool bushing from center block Step 9 Figure 9 With o ring pick gently remove pilot spool retaining o ring Replace if necessar...

Страница 30: ...the air valve end cap once the bolts have been removed Step 4 Figure 4 Remove air valve spool from air valve body by threading one air valve bolt into the end of the spool and gently sliding the spoo...

Страница 31: ...center section Step 9 Figure 9 With o ring pick gently remove the o ring from the opposite side of the center hole cut on the spool Gently remove the pilot spool from sleeve and inspect for nicks or g...

Страница 32: ...new seal With a new seal in hand place the two legs of the needle nose pliers inside the seal ring See Figure A Open the pliers as wide as the seal diameter will allow then with two fingers pull down...

Страница 33: ...expanded PTFE Gasket Kits P N 08 9501 99 and P8 polypropylene pumps come with kits for liquid chambers only P N 08 9500 99 Carefully prepare sealing surfaces by removing all debris and foreign matter...

Страница 34: ...N PUMP ENGINEERING LLC 32 WIL 10131 E 01 P8 PLASTIC R U B B E R T P E F I T T E D O R U LT R A F L E X T M F I T T E D E X P L O D E D V I E W S e c t i o n 8 E X P L O D E D V I E W P A R T L I S T I...

Страница 35: ...03 04 3890 03 20 Inner Piston 2 08 3700 01 08 3700 01 08 3700 01 08 3700 01 Inner Piston Ultra Flex 2 08 3761 01 08 3761 01 08 3761 01 08 3761 01 21 Diaphragm 2 22 Outer Piston 2 08 4550 21 500 08 455...

Страница 36: ...WILDEN PUMP ENGINEERING LLC 34 WIL 10131 E 01 S e c t i o n 8 E X P L O D E D V I E W P A R T L I S T I N G P8 PLASTIC P T F E F I T T E D E X P L O D E D V I E W...

Страница 37: ...4 3890 03 04 3890 03 04 3890 03 04 3890 03 20 Inner Piston 2 08 3750 01 08 3750 01 08 3750 01 08 3750 01 21 Diaphragm Back up 2 08 1060 51 08 1060 51 08 1060 51 08 1060 51 22 Diaphragm 2 08 1010 55 08...

Страница 38: ...WILDEN PUMP ENGINEERING LLC 36 WIL 10131 E 01 PX8 PLASTIC R u b b e r T P E U l t r a F l e x F i t t e d E X P L O D E D V I E W S e c t i o n 8 E X P L O D E D V I E W P A R T L I S T I N G...

Страница 39: ...3890 03 04 3890 03 21 Inner Piston 2 08 3700 01 08 3700 01 08 3700 01 08 3700 01 Inner Piston Ultra Flex 2 08 3761 01 08 3761 01 08 3761 01 08 3761 01 22 Diaphragm 2 23 Outer Piston 2 08 4550 21 500 0...

Страница 40: ...WILDEN PUMP ENGINEERING LLC 38 WIL 10131 E 01 S e c t i o n 8 E X P L O D E D V I E W P A R T L I S T I N G PX8 PLASTIC P T F E F i t t e d E X P L O D E D V I E W...

Страница 41: ...hragm Back up 2 08 1060 51 08 1060 51 08 1060 51 08 1060 51 23 Diaphragm 2 08 1010 55 08 1010 55 08 1010 55 08 1010 55 24 Outer Piston 2 08 4600 21 500 08 4600 21 500 08 4600 21 500 08 4600 21 500 25...

Страница 42: ...1020 54 N A N A 08 1080 54 Viton 08 1010 53 08 1020 53 N A N A 08 1080 53 Saniflex 08 1010 56 N A N A N A 08 1080 56 PTFE 08 1010 553 N A N A N A 08 1080 55 Neoprene Backup 08 1060 51 N A N A N A N A...

Страница 43: ...comes rst Failure due to normal wear misapplication or abuse is of course excluded from this warranty Since the use of Wilden pumps and parts is beyond our control we cannot guarantee the suitability...

Отзывы: