12. Technical specifications
Before commissioning, check that the voltage corresponds with the operating
voltage specified on the identification plate�
Dimensions
Height: 170mm
Width: 145mm
Depth: 200mm
Weight
AIRcon humidifier:
approx� 2�3kg
approx� 2�5kg
incl� delivery accessories
Classification
{
{
Unit (protection class cf� IEC 60601) Class II
{
{
Application parts of Type BF:
- Heated Breathing tube system
- Temperature probe
{
{
Protection category by enclosure IP22 (protected against
solid objects with diameter from 12�5mm, protected
against access with a finger, protection against falling
water drops if the enclosure is tilted up to 15°�)
Electrical
specifications
{
{
Operating voltage:
AIRcon Gen2 101200
220V~ to 240V~
AIRcon Gen2 101201
110V~ to 120V~
{
{
Mains frequency:
50Hz / 60Hz
{
{
Power consumption:
max� 315VA
{
{
Heating plate:
170W
{
{
Inspiratory tube heater:
22V~ , 30W
Operating data
{
{
Warm-up time:
max� 30min�
{
{
Recommended flow rate:
2 to 80l/min
{
{
Humidifier
>33mg/l
system output:
in the range of 2–60l/min
> 10mg/l
in the range of 2–80l/min
{
{
Maximum operating pressure: 200 mbar
1
{
{
Continuous noise:
< 50dB (1m)
{
{
Sound pressure level
of the alarm:
max� 65dB
{
{
Max� volume of water:
200ml