15
5.4 Connecting the
breathing tube system
The T- ensor of the temperature probe
must be in the opening of the angle
connector,
T sensor in the opening of the angle connector
and the sensor at the end of the cable
must be inserted into the patient-side
opening at the end of the breathing
tube system�
Both sensors must be firmly and
securely inserted in the respective
opening� The cable of the temperature
probe can be fixed in the corresponding
hook of the hose brackets�
The green coupling of the heating wire
power supply is always connected
to the green heating wire plug of the
inspiration hose�
Heating wire plug
The basic device AIRcon Gen2 is only
ready to operate if the inspiration
heating wire is connected�
If the breathing tube system used is
equipped with a heated expiration tube,
the white coupling of the heating wire
power supply is always connected with
the white heating filament plug of the
expiration tube�