GB
WIKA operating instructions temperature transmitter, model T24
21
2370242.09 08/2013 GB/D
WARNING!
In order to carry out any configuration, the positive and
negative connections must be disconnected from any
connected evaluation system.
7. T24 configuration software WIKA_TT
For installation please follow the instructions of the installation
instructions. A free-of-charge download of the up-to-date version of
the WIKA_TT software (suitable for Windows 3.xx/95/98/2000/ME/XP/
NT 4.0/7/VISTA) is available from www.wika.com.
7.1 Starting up the software
Start the software by double-clicking on the WIKA_TT icon.
7.2 Connection
Establishing connection to a model T24
temperature transmitter through the
menu item "Instrument"/"Download from
instrument". Displays the configuration data of
the connected T24.
Access to all operation-related functions and parameters such as:
■
Sensor type and connection
■
Measuring range and temperature unit
■
Error signalling in the event of sensor burnout
■
Test point Tag no. (M&C information)
■
Write protection and diagnostic functions
Connection is only possible to a single instrument at any
one time. During this, do not interrupt the communication
with the transmitter, since the data may not be correctly
read as a result of this.
6. Commissioning, operation / 7. T24 ...