11.1
Safety-related characteristic values for models T16.x-AI, T16.x-AC
Intrinsically safe connection values for the current loop (4 ... 20 mA)
Protection level Ex ia IIC/IIB/IIA, Ex ia IIIC or Ex ic IIC/IIB/IIA
Parameters
Models T16.x-AI, T16.x-AC
Models T16.x-AI
Gas hazardous application
Dust hazardous
application
Terminals
+ / -
+ / -
Voltage U
i
DC 30 V
DC 30 V
Current l
i
130 mA
130 mA
Power P
i
800 mW
750/650/550 mW
Effective internal capacitance C
i
18.4 nF
18.4 nF
Effective internal inductance L
i
800 µH
800 µH
Sensor circuit
Parameters
Models T16.x-AI
Model T16.x-AC
Ex ia IIC/IIB//IIA
Ex ia IIIC
Ex ic IIC/IIB//IIA
Terminals
1 - 2
1 - 2
Voltage U
o
DC 6.6 V
DC 6.6 V
Current I
o
4 mA
4 mA
Power P
o
10 mW
10 mW
Max. external
capacitance C
o
IIC
21 µF
1)
280 µF
1)
IIB IIIC
495 µF
1)
995 µF
1)
IIA
995 µF
1)
995 µF
1)
Max. external
inductance L
o
IIC
95 mH
95 mH
IIB IIIC
95 mH
95 mH
IIA
95 mH
95 mH
Characteristics
Linear
Comments:
U
o
: Maximum voltage of any conductor against the other three conductors
I
o
: Maximum output current for the least favourable connection of the internal current limiting resistors
P
o
: U
o
x I
o
divided by 4 (linear characteristic)
1) Internal L and C have already been considered
Due to distance requirements of the applied standards, the IS power and signal circuit as well
as the IS sensor circuit shall be considered as being galvanically connected to each other.
The electrical parameters of the head and rail mounting versions are identical.
Содержание T15 series
Страница 32: ...EN 14147933 02 10 2018 EN DE WIKA operating instructions model T16 32 Appendix 1 Control drawing CSA FM ...
Страница 33: ...EN 14147933 02 10 2018 EN DE WIKA operating instructions model T16 33 Appendix 1 Control drawing CSA FM ...
Страница 34: ...EN 14147933 02 10 2018 EN DE WIKA operating instructions model T16 34 Appendix 1 Control drawing CSA FM ...
Страница 35: ...EN 14147933 02 10 2018 EN DE WIKA operating instructions model T16 35 Appendix 1 Control drawing CSA FM ...
Страница 36: ...EN Appendix 2 EU declaration of conformity 14147933 02 10 2018 EN DE WIKA operating instructions model T16 36 ...
Страница 66: ...66 14147933 02 10 2018 EN DE WIKA operating instructions model T16 ...
Страница 67: ...67 14147933 02 10 2018 EN DE WIKA operating instructions model T16 ...