42
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
PEAKMAX?
<sp>n.nn<cr><lf>
Returns the maximum pressure since
peakreset was sent.
PEAKMIN?
<sp>n.nn<cr><lf>
Returns the minimum pressure since
peakreset was sent.
PEAKRESET
none
Resets the peak values.
PORT
nnnnnn
Sets the Ethernet port of the instrument.
PORT?
<sp>nnnnn<cr><lf>
Returns the Ethernet port of the instru-
ment.
PTYPE
Absolute, Gauge or
Differential
Sets the instrument pressure type - only
works if the optional barometric sensor is
installed.
PTYPE?
<sp>CCCCC<cr><lf>
Returns “Absolute” or “Gauge” for the
pressure type.
RANGEMAX?
<sp>XXXXXXX<cr><lf>
Returns the maximum range of the active
transducer in the current units.
RANGEMIN?
<sp>XXXXXXX<cr><lf>
Returns the minimum range of the active
transducer in the current units.
RATE?
<sp>XXXXXXX<cr><lf>
Returns the rate reading of the instrument
in current units/second.
RDECPT?
<sp>n<cr><lf>
Returns the number of rate decimal points
for the active channel. (see RESOLUTION)
RELEASE?
15 char string
Example:
Release? Test_stand_1
Returns: <sp>(yes or no),
CCC…CCC<cr><lf>
This command is used to release control
of the instrument in a multiple computer
environment. Yes if release is successful.
No if instrument is being controlled with
another computer. CCC… = name of con-
trolling computer or AVAILABLE.
See: ACQUIRE? and UNLOCK
RESOLUTION
n
Sets the number of significant digits.
See DECPT?
RESOLUTION?
<sp>n<cr><lf>
Returns the number of significant digits.
See DECPT?
RFILTER
value in %
Sets the % of the rate filter.
RFILTER?
<sp>n.nn<cr><lf>
Returns the rate filter.
RWINDOW
value in current units
Sets rate exponential filter window.