11
WIKA operating instructions pressure sensor, model O-10
EN
11613017.05 09/2019 EN/DE/FR/ES
3. Specifications
3.3 Output signals model O-10 (T)
Signal type
Signal
Current (2-wire)
4 ... 20 mA
Voltage (3-wire)
■
DC 0.5 ... 4.5 V
■
DC 0 ... 5 V
■
DC 1 ... 5 V
■
DC 0 ... 10 V
Ratiometric (3-wire)
DC 0.5 ... 4.5 V
Load in Ω
■
Current output (2-wire):
≤ (power supply - 8 V) / 0.02 A
■
Voltage output (3-wire):
> maximum output signal / 1 mA
■
Ratiometric output signal (3-wire):
> 4.5 kΩ
3.4 Output signals model O-10 (5)
Signal type
Signal
Current (2-wire)
4 ... 20 mA
Voltage (3-wire)
■
DC 0.5 ... 4.5 V
■
DC 1 ... 5 V
Ratiometric (3-wire)
DC 0.5 ... 4.5 V
Load in Ω
■
Current output (2-wire):
≤ (power supply - 8 V) / 0.02 A
■
Voltage output (3-wire):
> maximum output signal / 1 mA
■
Ratiometric output signal (3-wire):
> 4.5 kΩ