EN
64
WIKA operating instructions, model CTR3000
14191761.01 05/2019 EN/DE
Up to 64 channels may be added with the addition of external CTS5000 multiplexers.
Two multiplexer models are available:
■
Model CTS5000-8 with 8 channels
■
Model CTS5000-16 with 16 channels
The multiplexers use an expansion connector on the rear of the instrument. Channels PRT1, PRT2, TC1 and TC2 operate
independently of the multiplexers.
Multiplexer channels are selected by their channel numbers:
Multiplexer
CTS5000-8
Channel number
CTS5000-16
Channel number
1
1.1 to 1.8
1.1 to 1.16
2
2.1 to 2.8
2.1 to 2.16
3
3.1 to 3.8
3.1 to 3.16
4
4.1 to 4.8
4.1 to 4.16
CTS5000-16’s and CTS5000-8’s may be mixed in any order.
6.7.2 Multiplexer model CTS3000
Set up
Multiplexers are detected at power-on and are daisy-chained. Addresses are assigned in order, so the first multiplexer
connected to the CTR3000 will start with the lowest address.
An expansion port connector is provided on the rear panel at the CTR3000 itself. Here you have to connect the CTS expansion
and TC expansion port with the delivered cables.
The CTS connector is a (1:1) 15-way socket for signal, power and data connection to the CTS3000. The TC expansion cable is
a 3-pin TC connector.
USB Device
Ethernet
RS-232 / IEEE Port
TC Expansion
Power Input
CTS Expansion
CTS Expansion
TC Expansion
IN
IN
OUT
OUT
6. Commissioning, operation