16
WIKA operating instructions, model CTD9100-1100
14030725.02 08/2019 EN/DE
EN
Function
Meaning
Units °C/°F/K
Setting of the unit, which should be displayed
DEF.PAR. ON/OFF Default Parameter
The controller can handle default and customized settings of P.B./T.I./T.D.
■
OFF: customised settings
■
ON: default settings
The controller is adapted to the calibration ex-factory. If there are any change wishes,
please contact the manufacturer.
Do not change these settings!
KEY
Code for the functions in the third level (Default 2)
7.2.4 Third level - Operation settings
Following functions can only be reached, if the
„KEY“
in the second level is entered over the buttons
[▲]
/
[▼]
is correct and
then the buttons
[M]
and
[▲]
have to be pressed simultaneously. Back over simultaneous pressing of the buttons
[M]
and
[▲]
or waiting 20 s.
Function
Meaning
Access key
Access code
(Default = 2)
Baud rate
Velocity of data transfer
(Default = 9600)
Address
Communication address
(Values from 1 to 32 possible)
S/N
Serial number of the instrument
(not editable)
Board S/N
Serial number of the board
(not editable)
MAX. SET.
Maximal temperature set point
(not editable)
MIN. SET.
Minimal temperature set point
(not editable)
WAIT 0/1
Setting “0”: Calibrator starts to control the set point after switching it on
Setting “1”: Calibrator starts to control after switching it on, when the tester has confirmed
REV. SOFTWARE
Firmware Version
SENSOR TYPE
Type of internal reference sensor
STAB: +/-0.3 °C
Symbol ÷ appears on the display, if the temperature is stable < +/- 0, 3 °C over 6 minutes.
7. Operating of the calibrator