20
WIKA operating instructions, reference pressure sensor, model CPT62I0
EN
14225194.01 12/2018 EN/DE/FR/ES
Reference pressure sensor model CPT62I0
Material
Wetted parts
Stainless steel or Elgiloy
®
, (> 25 bar (360 psi) additionally with NBR
sealing)
Flush diaphragm version: Stainless steel {Hastelloy C4}; O-ring:
NBR {FKM/FPM or EPDM}
Internal transmission
medium
Synthetic oil (only for measuring ranges to 16 bar (250 psi) or flush
diaphragm)
{Halocarbon oil for oxygen applications}; {Listed by FDA for food
industry}
Sensor data
Accuracy
2)
≤ 0.2 % of span (at reference conditions)
3)
Compensated range
0 ... 80 °C (0 ... 176 °F)
Mean temperature
coefficient
≤ 0.2 % of span/10 K (outside the reference conditions)
Permissible ambient conditions
Medium temperature
1)
-20 ... +50 °C (-4 ... +122 °F)
Operating temperature
-20 ... +50 °C (-4 ... +122 °F)
Storage temperature
-40 ... +80 °C (-40 ... +176 °F)
Relative humidity
0 ... 95 % r. h. (non-condensing)
Case
Material
Stainless steel
Connection to the
CPH62I0
via 1.1 m (3.3 ft) sensor connection cable (plug-and-play);
Optional: Extension cable approx. 3.8 m (12.5 ft), overall cable
length approx. 5 m (16.4 ft)
Ingress protection
IP65
IP67 when connected
Dimensions
See technical drawing
Weight
approx. 220 g (0.49 lbs)
{} Items in curved brackets are optional extras for an additional price.
1) As an oxygen version, a flush diaphragm model is not available. In an oxygen version, the model CPT62I0 is only available in
overpressure ranges ≥ 0.25 bar (≥ 0.4 psi), with medium temperatures between -10 … +50 °C (14 … 122 °F) and using stainless
steel or Elgiloy
®
wetted parts.
2) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following
factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability,
influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point correction.
3) Reference conditions: 15 ... 25 °C (59 ... 77 °F)
9. Specifications