
21
WIKA operating instructions, model CPH65I0
14061681.02 03/2013 GB/D
GB
6. Commissioning, operation
6. Commissioning, operation
As soon as the model CPH65I0 hand-held pressure calibrator is turned on, by pressing
the
ON/OFF
key, it will go through a short self-test routine. During this routine, the
display shows the current firmware version, auto-shutdown status and the measuring
range of the internal pressure sensor. The calibrator requires a maximum of 5 minutes
warm-up to reach its specified accuracy. Large changes in ambient temperature may
make a longer warm-up period necessary. The pressure display of the calibrator should
be zeroed before starting the pressure calibration. See chapter 6.3 "Use of the
"ZERO"
function".
6.1 Calibrator display
The calibrator's display consists of two sectors: The menu bar (located along the
bottom of the screen) is used to access the menu system. The main display (the rest of
the display) consists of up to three sub-sectors for the measurement activities. These
sub-sectors will be referred to as the
UPPER
,
MIDDLE
and
LOWER
displays in this
document.
Figure 1 shows the location of the different display fields, which are described in the
following table.
1)
Pressure units
Indication of the pressure unit (selectable from 16 pressure units)
2)
Units
Indication of the measuring unit
3)
Display of the span
Indication of 4 ... 20 mA span (only with mA measurement)
4)
Menu list
5)
Primary parameters
Indication of the current measured parameters
Figure 1: Display
1
2
3
4
5
Содержание CPH65I0
Страница 33: ...33 WIKA operating instructions model CPH65I0 14061681 02 03 2013 GB D GB 6 Commissioning operation...
Страница 50: ...50 WIKA operating instructions model CPH65I0 14061681 02 03 2013 GB D GB...
Страница 80: ...80 WIKA Betriebsanleitung Typ CPH65I0 14061681 02 03 2013 GB D D 6 Inbetriebnahme Betrieb 6 6 Men struktur...
Страница 81: ...81 WIKA Betriebsanleitung Typ CPH65I0 14061681 02 03 2013 GB D D 6 Inbetriebnahme Betrieb...
Страница 97: ...97 WIKA Betriebsanleitung Typ CPH65I0 14061681 02 03 2013 GB D D Anlage 1 Konformit tserkl rung Typ CPH65I0...