6.5.2.1 Starting logger recording
1. Press the
[STORE/QUIT]
button for 2 seconds.
⇒
The display shows “
Lo66
rvn
”:
2. Press the
[STORE/QUIT]
button once more.
⇒
Recording starts.
With each recording, the display will briefly show “
St.XXXX
”. Here, XXXX is the number of
the data set 1 ... 4,000 or 1 ... 10,000.
The following displays may appear for logger recording:
Lo66
rvn
Displayed when logger recording is to be started.
Lo66
CLr
Function “Delete” will be selected
Lo66
FvLL
Shows in the event the logger memory is full. The recording will be
stopped automatically.
Lo
.
Po
=
on
Appears with the low-power logger function.
As soon as the logger memory is full, the instrument switches itself
off.
6.5.2.2 Stopping the logger recording
1. Press the
[STORE/QUIT]
button briefly.
⇒
A confirmation prompt then appears:
2. With the
[▼/MIN]
or
[▲/MAX]
button, select the desired function.
The following functions can be selected:
SToP
yes
Stop recording.
SToP
no
Do not stop recording (cancel the process).
3. With the
[STORE/QUIT]
button, confirm the selection.
4. Press the
[SET/MENU]
button.
⇒
Return to main menu.
If, during a cyclic data logging, the measuring instrument is switched off, you
will automatically be asked whether the data logging should be stopped. The
instrument can only be switched off after the recording has been stopped.
The auto power-off function is deactivated during recording!
6. Commissioning, operation