WIKA operating instructions models TR10-0, TC10-0 (FM, CSA)
EN
9. Maintenance, cleaning and recalibration
For contact details, please see chapter 1 “General information” or the back
page of the operating instructions.
9.1 Maintenance
The thermometers described here are maintenance-free.
Repairs must only be carried out by the manufacturer.
9.2 Cleaning
CAUTION!
Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and
the environment. Residual media in the dismounted instrument can result in a
risk to persons, the environment and equipment.
▶
Carry out the cleaning process as described below.
1. Prior to cleaning the electrical connections, disconnect them properly.
2. Clean the instrument with a moist cloth.
3. Electrical connections must not come into contact with moisture.
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any pointed and hard objects for cleaning.
4. Wash or clean the dismounted instrument, in order to protect persons and the environ-
ment from exposure to residual media.
Residual media in the dismounted instrument can result in a risk to persons, the environ-
ment and equipment. Take sufficient precautionary measures.
9.3 Calibration, recalibration
It is recommended that the measuring insert is recalibrated at regular intervals (resistance
thermometers: approx. 24 months, thermocouples: approx. 12 months). This period can
reduce, depending on the particular application. The calibration can be carried out by the
manufacturer, as well as on-site by qualified technical staff with calibration instruments.
For calibration, the measuring insert is removed from the thermometer. The minimum length
(metal part of the probe) for carrying out a measurement accuracy test 3.1 or DAkkS is
100 mm [3.94 in].
Calibration of shorter lengths on request.
9. Maintenance, cleaning and recalibration