The UNIQUE™ series hearing aids are intended for use in the electromagnet-
ic environment specified below. The customer or the user of a UNIQUE™ ser-
ies hearing aid should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic dis-
charge (ESD)
IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV con-
tact
± 8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidi-
ty should be at least
30%.
Electrical fast tran-
sients/burst
IEC
61000-4-4
± 2 kV for power line
supplies
± 1 kV for input/out-
put lines
Not
applicable *)
Not
applicable *)
Not
applicable *)
34