55
Immunity
Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Voltage
dips, short
interruptions
and voltage
variations on
power sup-
ply input
lines
IEC
61000-4-11
<5 %
U
T
(>95 % dip
in
U
T
) for
0.5 cycle
40 %
U
T
(60 % dip
in
U
T
) for 5
cycles
70 %
U
T
(30 % dip
in
U
T
) for 25
cycles
<5 %
U
T
(>95 % dip
in
U
T
) for 5 s
Not
applicable *)
Not
applicable *)
Power fre-
quency
(50/60 Hz)
magnetic
field
IEC
61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at lev-
els characteristic
of a typical loca-
tion in a typical
commercial or
hospital environ-
ment
NOTE U
T
is the a.c. mains voltage prior to the application of the
test level.
*) Battery powered equipment