82
Electromagnetic immunity
The DREAM™ series hearing aids are intended for use in the electro-
magnetic environment specified below . The customer or the user of a
DREAM™ series hearing aid should assure that it is used in such an envi-
ronment .
Immunity
Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment – guid-
ance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
± 6 kV con-
tact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete or
ceramic tile . If floors
are covered with syn-
thetic material, the
relative humidity
should be at least
30% .