21
2. To ensure the accuracy of the measurement, please wait at least 30
seconds after 5 consecutive measurements.
12. Statement of EMC
This product meets YY0505-2012 requirements, but all types of electronic
devices may cause electromagnetic interference to other devices through air.
The following requirements should be strictly observed in the use of the
product, otherwise it may cause electromagnetic interference to other
equipment or reduce the anti-electromagnetic interference ability of the
product, bring safety safety risks or loss of product performance.
1) Portable and mobile radio frequency communication equipment may
affect the description of medical electrical equipment: portable and
mobile radio frequency communication equipment may affect the normal
operation of this product. Portable and mobile radio frequency
communication equipment and this product should meet a certain space
distance, see Table 4.
2) Warning: Except for the accessories sold by the manufacturer of the
product as spare parts for internal components, the use of additional
accessories may lead to an increase in equipment or system emissions or
a decrease in immunity.
3) Warning: This product should not be used close to or stacked with other
equipment. If it must be used close to or stacked, it should be observed
and verified to operate normally under the configuration in which it is
used.
4) Warning: The operation of this product below the minimum amplitude or
minimum value described in this manual may lead to inaccurate
consequences.
5) This product meets the requirements of EMC of medical devices. In the
EMI environment in Table 2-4, it should meet the basic performance
requirements.
6) Basic performance
a) Requirements: the equipment should be able to display normally, the
normal function of the button should not fail, the maximum allowable
error in the temperature range of 35.0℃~42.0℃ is ±0.2℃.
b) Test: the infrared thermometer is switched to the object temperature
mode, and the blackbody temperature is set to 35.0℃ . Before the test,
the thermometer should be stable for at least 30min under the laboratory
reference condition, and the maximum allowable error of the measured
reading should meet the requirements.
7) This product statement meets the requirements of Table 1-4 below.
Table 1
Guide and Manufacturer's Statement –electromagnetic Launch
This product is expected to be used in the electromagnetic environment
specified below, and the purchaser or user shall ensure that it is used in such
electromagnetic environment
Emission Test
Compliance
electromagnetic Environment - Guide
Radio-frequency
emission GB 4824
Group 1
This product uses RF energy only for its
internal functions. Therefore, its RF