3

Ÿ

On the Little Splint™, the dual thermostat control is  

 

set with a high and low setting. The high setting is  

 

196°F (91°C) and the low setting is 109°F (43°C).  

 

Allow the unit to reach temperature. 

Ÿ

Fill the unit with water to 1" deep. Plug the unit into a  

 

properly grounded outlet having a ground fault circuit  

 

interrupter either in the receptacle or somewhere in  

 

the immediate power line.

Ÿ

Place the Little Splint™ on a secure platform, i.e., a  

 

table or big cart. 

Ÿ

With 1" of cold water and with the cover closed,  

 

operating temperature should be reached in    

 

approximately 30 minutes. To reduce the warm up  

 

time for the Little Splint, you may use hot water.

Ÿ

Note: We recommend the use of distilled water to  

 

reduce the build-up of mineral deposits and dirt.

It may be necessary or you may desire to return the 
Little Splint™ unit for service. In that case, we would 

®

appreciate you contacting Whitehall Manufacturing  
directly. We will make arrangements for the unit to be 
returned to our production location. When the unit is 
received by us, an estimate of repairs will be made and 
you will be contacted. 

We prefer warranty repair or 

any other repair be arranged directly with Whitehall 
Manufacturing. If you are not satisfied with our 
service, then and only then, contact your dealer.

CHANGING THE FUSE

Turn the Little Splint off via the controls on the front panel 
and then unplug the unit. Allow the unit and the water to 
cool before draining the water from the unit. Next, 
position the unit to allow access to the rear of the splint 
pan. Locate the fuse holder caps with the use of a 
"common" blade screwdriver or fingers and rotate  

counterclockwise  

approximately 90° or until the cap 

becomes loose. Remove the fuse holder cap, exposing 
the ceramic fuse. Remove the blown fuse.

Once the old fuse has been removed, replace it with a 
new fuse of the same kind and amperage. Reinstall the 
fuse holder cap by depressing and rotating 

clockwise

 

approximately 90° and ensure that the fuse holder caps 
are secured. Refill the Little Splint™ with at least 1" of 
water and then plug the unit back into a power outlet. 
Turn the unit on via front control panel. Allow the water to 
warm up to the operating temperature and use as 
required.

TO CHANGE THE TEMPERATURE SETTING

The temperature is varied by turning the thermostat knob 
clockwise for an increase and counterclockwise for a 
decrease in temperature. The temperature on low setting 
should approximate 109°F (43°C) and on high 196°F 
(91°C). 

REMOVAL OF WATER

Cool the bath and tip the unit and drain the water into a 
bucket, sink or drain. 

Again, disinfection procedures are different from facility 
to facility, and we have found that alcohol is still the best 
cleaning and solution. Bleaches and iodine should not 
be used.

Use of bleach or iodine will result in pin holes in the 
stainless steel tank which will leak and void the warranty.

Cleaning of the pan can best be accomplished with a 
very mild, non-scratching household cleaner and a very 
low abrasion plastic scouring pad lightly applied. On a 
stainless steel tank, never use a metal scouring pad as it 
is too abrasive and may leave iron deposits on the 
stainless, which will rust.

Содержание LITTLE Splint LS-1

Страница 1: ...ATION AND CARE OF LITTLE Compact Splint Pan Model LS 1 6900 191 001_REV_A 800 782 7706 626 968 6681 www whitehallmfg com P O Box 3527 City of Industry CA 91744 0527 U S A Division of Acorn Engineering...

Страница 2: ...iqu l EFFECTUER LA MISE LA TERRE de votre appareil DE FA ON APPROPRI E Brancher l appareil dans une prise disjoncteur diff rentiel de fuite la terre DDFT ce qui offre une protection au cas o l eau et...

Страница 3: ...rature Setting 3 Changing the Fuse 3 Repair of Unit 3 Parts Diagram Parts List and Schematic 4 3 WHITEHALL MANUFACTURING P O BOX 3527 City of Industry CA 91744 0527 U S A Phone 800 782 7706 626 968 66...

Страница 4: ...id is functionally designed to open completely with just one hand leaving the other hand free The unit is built out of stainless steel for long lasting durability The Little Splint s tank interior is...

Страница 5: ...ont panel and then unplug the unit Allow the unit and the water to cool before draining the water from the unit Next position the unit to allow access to the rear of the splint pan Locate the fuse hol...

Страница 6: ...6505 056 000 Power Cord 6 6903 147 000 Thermostat Manual Reset 7 6900 102 000 Clamp Steel Cable 8 0302 011 000 6 32 UNC Stainless Steel Hex Nut 9 0322 003 000 6 Plated Star Lock Washer External 1 6505...

Страница 7: ...have been damaged by other causes such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy of the Buyer and the sole...

Отзывы: