5

Warnings 

Treatments:  

  Whirlpools are intended for use by a medical professional only, e.g., Physiotherapist, 

Physical Therapist, Rehabilitation Therapist, Athletic Trainer, Podiatrist. 

  During treatment, the medical professional must stay with patient at all times.  

  For Whirlpools that support that the entire mass of the patient, the maximum allowable 

patient mass is 135 kg (298 lbs). 

  Ensure water in tub is no greater than 43.3ºC (110ºF).  Higher temperatures may result 

in injury to patient.  

 

Treatment should terminate if the patient experiences dizziness, light-headedness, 
shortness of breath, or nausea

 

To avoid patient cross-infection,

 drain and clean hydrotherapy equipment after each 

patient’s use, and disinfect equipment surfaces and components by using an EPA-
registered product in accordance with the manufacturer’s instructions. 

Operation: 

  Use of this equipment adjacent to or stacked with other equipment should be avoided 

because it could result in improper operation. If such use is necessary, this equipment 
and the other equipment should be observed to verify that they are operating normally. 

  Use of accessories and cables other than those specified or provided by the 

manufacturer of this equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and result in improper 
operation. 

  Use only Whitehall-approved items on the Whirlpool. 

  Portable RF communications equipment (including peripherals such as antenna cables 

and external antennas) should be used no closer than 30 cm (12 inches) to any part of 
the Whirlpool, including cables specified by the manufacturer. Otherwise, degradation 
of the performance of this equipment could result. 

  If there is visible damage to the 

Whirlpool

 or packaging upon receipt, report this 

damage to shipping company and Whitehall.  Do not use the Whirlpool if damage to it 
is suspected.  

  Operate with the bottom of the turbine submerged in at least 15.2 cm (6 inches) of 

water. 

  Keep hair and loose items (gown straps, gauze, etc.) out of the tub.  These items can 

be pulled into the pump and cause injury or damage to the turbine.  

  Use the Whirlpool only for the purpose described in this manual. 

  Position the Whirlpool so that the power cord is easily accessible. 

Содержание E-15-M

Страница 1: ...H P 15 MU P 22 M Extremity E 15 M E 15 MH E 15 MU E 22 M E 22 MU E 27 M E 45 M Hi Boy H 60 M H 75 M H 90 M H 105 M Lo Boy L 75 M L 90 M L 105 M Sports S 85 M S 90 M S 110 M Slant Back SB 100 M P O Box...

Страница 2: ...Manufacturing a division of Acorn Engineering and a member of Morris Group International Any reproduction dissemination or distribution of this document in whole or in part without prior written auth...

Страница 3: ...e 7 Contraindications 7 User Skills 7 Device Description 8 Options and Accessories 9 Facility Requirements 11 Instructions 12 Assembly Instructions 12 Operating Instructions 12 Draining Instructions 1...

Страница 4: ...o drop off the waste equipment for recycling contact your local waste collection authority General Warning or Caution Indicates personal injury or damage to the equipment is possible Electric Shock In...

Страница 5: ...improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally Use of accessories and cables other than those specifi...

Страница 6: ...cuit Interrupter GFCI medical grade receptacle No parts of the Whirlpool including the power supply cord are repairable or replaceable by service personnel No modification of the Whirlpool is allowed...

Страница 7: ...n which hydrotherapy is contraindicated Active infection anywhere in the body Open wounds Loss of temperature sensation Hydrophobia Heat or cold intolerance Poor balance Kidney failure Safety Effectiv...

Страница 8: ...water from the pump to a drain hose mounted at the top of the turbine by an optional drain pump located below the tub or by opening up the drain valve at the bottom of the tub Four casters all of whic...

Страница 9: ...he color of the powder coat Powder Coat Paint All except Slant Back LOGO Logo selected by customer All except Slant Back The following accessories are available Description Used on Models Instructions...

Страница 10: ...maintenance see the Disinfecting and Cleaning Instructions section TTS2 Tank Top Seat for 610 mm 24 inch wide tubs H 105 M S 90 M S 110 M L 75 M L 90 M L 105 M ARS Arm Rest Support All Hang support o...

Страница 11: ...licensed electrician Environmental conditions within those specified in the Specification section of this manual NOTE The emissions characteristics of this equipment make it suitable for use in indus...

Страница 12: ...rough hose if discharge nozzle 5 is lower than water level 4 Turn the pressure control valve handle 11 fully clockwise then fully counterclockwise to completely lubricate the plunger assembly 5 For Wh...

Страница 13: ...p 12 to the bottom of the tub c Place discharge nozzle 5 into the sink d Turn the turbine power switch 13 on e When water stops discharging from the nozzle turn off the switch f Drain remaining water...

Страница 14: ...at least 10 2 cm 4 inches of water 2 Place the stem tip of a calibrated thermometer next to the tip of the Whirlpool thermometer 3 Turn on the turbine assembly 4 If the reading on the Whirlpool thermo...

Страница 15: ...ermometer and turbine shall be disinfected and cleaned after each patient Disinfecting and Cleaning the Tub and Thermometer 1 Drain the tub following the draining instructions above 2 Verify that the...

Страница 16: ...ause Remedy Loss in pressure Foreign material caught in pump inlet holes 1 Turn off and unplug turbine assembly 2 Remove foreign materials from bottom of pump Turbine pump is not submerged fully in wa...

Страница 17: ...A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 www whitehallmfg com 17 4 Replace the fuses with 6525 608 000 for the 125V turbine and 6525 612 000 for the 230V turbine or equivalent 5 Replace the...

Страница 18: ...90 L 105 344 9 91 1 398 6 105 3 445 9 117 8 Water flow rate 0 56 8 L min 0 15 US gallons min Electrical Input voltage 115 230 V 230 V models are for export only Line Frequency 50 60 Hz Rated Current...

Страница 19: ...527 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 www whitehallmfg com 19 Transportation and Storage 0 60 C 32 140 F Relative humidity 5 95 non condensing Atmospheric Pressure 70 106 kPa Ther...

Страница 20: ...such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy of the Buyer and the sole remedy of Whitehall under this war...

Отзывы: