INSTALLATION, OPERATION & MAINTENANCE MANUAL 8194

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

OPERATION AND MAINTENANCE

1

1

12

2

2

13

6

8

11

14

9

4

3

5

10

7

Operation

Maintenance

Shut-Off Ball Valve

Ÿ

Exercise this valve regularly to eliminate the valve from locking in position 

due to corrosion or hardness of the water supply.

Cleaning

Use only cleaning agents intended for metals such as stainless steel, copper 
and brass. Do not use any cleaning agents that may cause and allergic 
reaction to these metals. Warranty will not cover any deterioration of the 
products do to neglectful cleaning processes.

4.
5.
6.

8.
9.

11.
12.

1/2" NCT SUPPLY INLETS
(CAPPED FOR SHIPPING)

(4) MOUNTING TABS

SUPPLY SHUT-OFF BALL VALVES

3/4" MALE HOSE CONNECTION
(SHOWN CAPPED)

TRAP PRIMER

SECOND BACKFLOW PREVENTER VALVE

FIRST BACKFLOW PREVENTER VALVE

2-3/8" O.D. WASTE CONNECTION

DISCHARGE HOSE BRACKET

13.

THERMOSTATIC MIXING VALVE

TEMPERATURE GAUGE

TRAP PRIMER OUTLET

Parts List

7. FOLD-DOWN DRIP TRAY

10.

BACKFLOW PREVENTER

14.

1.

2.
3.

R

Backflow Preventer

Ÿ

This product must be tested and certified prior to service. Certification must 

be done after final connection and prior to service.

Ÿ

This product must be tested and certified periodically in compliance with local 

codes, but at least once per year or more as service conditions warrant.

Ÿ

Refer to WATTS   009QT 1/2” BRONZE REDUCED PRESSURE ZONE 

manual for all testing and maintenance guides. (

)

https://www.watts.com

Trap Primer (optional)

Ÿ

This product may need regular cleaning, maintenance or replacement of 

the screen due to sediment buildup.

Ÿ

Refer to JAY R. SMITH MFG. CO.   FIGURE NO. 2694 trouble shooting 

and maintenance guides. (

)

https://www.jrsmith.com

R

IMPORTANT: Owner/ Installer is responsible for checking applicable state and local codes to ensure that the Backflow Preventer 
is installed in compliance. This includes testing requirements and proper installation height above the ground. Thermostatic 
Mixing Valve is factory pre-set at 77 F. Verify requirements before altering temperature settings since they may affect patient 
health and safety. Thoroughly test outlet water temperature. 

Ÿ

Fold the drip tray (7) down to help eliminate any spills onto cabinetry or 

finished surfaces from the dialysis machine discharge line. Fold the drip tray to 
the vertical position prior to closing the access door.

Ÿ

Ensure you have a proper connection from the dialysis machine supply hose 

and the male hose thread connection on the 8194, prior to opening the second 
valve on the backflow preventer.

Ÿ

The first ball valve on the backflow preventer (9) is normally open. This valve 

is used for isolation during testing and certification of the backflow preventer.

Ÿ

Thermostatic mixing valve is preset to 77 F and must be open and connected 

to supply and drainage hoses. If required, adjust the thermostatic mixing valve 
(3) to preferred temperature. See parts breakdown for instructions.

Ÿ

The second ball valve on the backflow preventer (10) is the on/off valve. This 

valve must be open during dialysis service and then closed prior to removal of 
the supply hose to the dialysis machine.

Ÿ

The supply shut-off ball valves (2) is normally open. This valve is intended to 

isolate the system for routine maintenance to the trap primer and backflow 
prevention device.

Ÿ

The dialysis machine discharge hose should be placed in the supplied hose 

bracket (8) to maintain the proper air gap from sanitary drain.