Whip Mix ProCal Скачать руководство пользователя страница 2

ProCal

 Operating Instructions 

Product Overview

The Pro Cal Calibration Kit is a professional grade 
instrumentation system designed to rapidly monitor the 
calibration status of your furnace, and automatically 
recalibrate the furnace to factory specifications.

It is recommended that you read this manual thoroughly 
before using the kit as there are valuable examples of 
procedures and decisions you will want to be acquainted 
with in order to maximize the benefits you’ve invested in.

What You Can Do With It

QC = Quick Check (Pro 200 Series). For FireLite refer to 
page 9.

The Quick Check feature allows you to rapidly spot check 
the need for calibration. Simply place the Special Firing 
Tray on the Lift Table, connect the Tray to the Instrument 
Box and the Instrument Box to the furnace calibration 
port, and dial in the desired temperature you want to 
verify. The LCD readout on the ProCal box should agree 
with the furnace display to within three degrees Celsius. If 
not, you can choose one of several options:

1.  If you check several temperatures and only one 

seems to be different, you can go into the firing 
programs which 

have hold temperatures at that 

value

 and use the 

Program Level Fusing Adjust

 to 

fine tune just those programs.

2.  If you check several temperatures and 

they all either 

overfire or underfire

, you can go into the 

Special 

Functions

 menu and select 

High Fusing Adjust for 

temperatures over 800°C

 or 

Low Fusing Adjust for 

temperatures under 800°C

 and enter the average 

correction value needed.

3.  You can do a fully automated calibration.

Full Recalibration (Pro 200 Series Only)

The full recalibration mode requires several hours to 
complete, so you’ll want to reserve it for times when the 
temperature is 

more than three degrees 

from nominal. 

Why does it take so long? To get a valid calibration 
which will most closely match the conditions in which it 
is used every day the furnace must be fully heat soaked. 
Secondly, the automatic calibration program requires 
one pass at three separate temperatures to build the 
calibration file. Thirdly, a second complete pass is 
necessary to confirm that the calibration pass is valid. 
This is exactly the same rigorous process that your 
furnace went through before it left the factory originally.

Why Do Furnaces Require Periodic 

Temperature Recalibration?

There are several reasons why furnaces require periodic 
QC checking and recalibration:

Effluents

 given off by the materials fired in the furnace 

can deposit on the platinum/palladium thermocouple and 

in some cases, chemically react with the noble metals, 
thereby affecting the temperature sensed.

Oxide layers

 can build up on the thermocouple 

through many cycles of vacuum, then exposure to 
air and chemically react with the noble metals in the 
thermocouple.

Prolonged high heat exposure

 can eventually crystallize 

the noble metals making them brittle and inhibiting the 
buildup of charge on each leg. The millivolts given off by 
a thermocouple to register temperature accumulate all 
along the heat exposed legs of the thermocouple, not just 
at the tip.

The heater wire also becomes brittle and scaly along its 
surface after extended usage. This can affect the infrared 
spectrum irradiated by the heater and redirect the angle 
which the photons emitted by the heater such that the 
ones striking the thermocouple are more diffused and 
therefore create less charge buildup in the thermocouple, 
affecting the calibration.

How It Works - What’s it doing in there?

The thermocouple in the Special Firing Tray monitors the 
furnace activity and reports its accumulated charge back to 
the Instrument Box. The Instrument Box filters the electrical 
noise inherent with very low level signals in the microvolt to 
millivolt range. The cleaned-up signal is then sent to a pre-
amplifier with a built in digital filter to further boost the signal 
and clean it up some more. Finally, the signal is presented 
to a 24 bit Analog to Digital Converter (ADC) for conversion 
into digital ones and zeros to be transmitted to the furnace 
computer. There is also a local “ambient” temperature 
sensor inside the Instrument Box which monitors the “cold 
junction” temperature at the point where the thermocouple 
alloys transition into the electronics. This cold junction 
temperature sensor output is automatically used by 

the 

computer inside the Instrument Box 

to correct any 

measurement errors as the reference signal from the 
Special Firing Tray makes its way back to the furnace.

Once the digital reference value from the Special Firing 
Tray is received back in the furnace computer its value is 
compared to the internal furnace thermocouple and the 
sign (relative over-firing or under-firing) and the magnitude 
of the difference are stored permanently in the furnace 
computer.

After the calibration process is complete, every time 
the furnace temperature is commanded to a specific 
temperature, the furnace computer looks at the target 
temperature, determines which calibration value is 
appropriate for that target value, and either adds or 
subtracts the stored calibration correction value in order to 
produce the accurate temperature desired, all in real-time.

1

Содержание ProCal

Страница 1: ...ProCal User Manual Whip Mix Corporation 361 Farmington Ave P O Box 17183 Louisville KY 40217 0183 USA 502 637 1451 800 626 5651 Fax 502 634 4512 www whipmix com ...

Страница 2: ...on the platinum palladium thermocouple and in some cases chemically react with the noble metals thereby affecting the temperature sensed Oxide layers can build up on the thermocouple through many cycles of vacuum then exposure to air and chemically react with the noble metals in the thermocouple Prolonged high heat exposure can eventually crystallize the noble metals making them brittle and inhibi...

Страница 3: ...epends upon these two components always being used together If more than one Kit is on site never use the Special Firing Tray from one Kit with the Instrument box of another Kit or inaccurate calibrations will result Since High Temperatures are present in the Furnace for extended periods of time it is NOT RECOMMENDED that Full Calibrations be attempted unless someone is present in the vicinity of ...

Страница 4: ...wire umbilical coming out of the edge of the Special Firing Tray is facing the front of the furnace Setup for Either Quick Check or Full Calibration Pro 200 only Use the Furnace Up Arrow button to raise the table until it is fully closed and the wire umbilical is protruding out of the front center of the muffle NOTE IT IS IMPOSSIBLE TO PULL VACUUM WITH THE SPECIAL FIRING TRAY IN PLACE BECAUSE OF T...

Страница 5: ...ting mode Allow Stabilization Time It is suggested that a period of at least 15 20 minutes is allowed for stabilization time before manually recording the value on the Instrument box LCD display to insure a stable reading What to do with the Results It is suggested that you survey three temperatures minimum to look for overall trends and to help you decide which type of correction method best fits...

Страница 6: ...Instrument Box is always the reference point not the furnace So the furnace is running two degrees HIGHER than the Instrument Box in other words the furnace is OVERFIRING by two degrees You would answer Yes to the High Overfire question in screen 3 above Yes is the default already displayed by pressing the Selector Knob OR pressing the Next button and then in screen 4 above you would roll the Sele...

Страница 7: ...Select Language Cal Low Fusing Cal High Fusing 11 High OverFire Yes 12 High Temp Adjustment 0C 0C 50C 13 Main Menu Run Program Change Program Special Functions You will be returned to Main Menu How to Enter An Overfire Correction Using The Program Level Fusing Adjust Pro 200 only Roll Selector Knob to item push Knob to select Press Knob to Skip Roll Knob to item push Knob to select Press Knob to S...

Страница 8: ... Preheat version of the Full Calibration Run be used for best accuracy The difference in Run Time is 27 but even the shorter version requires 2 15 hours versus the cool muffle version Run Time of 4 25 hours How to Launch the Calibration Run See the Physical Setup Section Described Previously to configure the Kit for a Run then From the Pro 200 Main Menu Select Special Functions Main Menu Special F...

Страница 9: ... results for 850 C 1000 C and 1150 C are on the confirmation screen shown above Anytime Thereafter you can retrieve the information by doing the following Main Menu Special Functions Calibrate Furnace Password 292 Profile Results 850 C 849 C 1000 C 999 C 1150 C 1151 C Pro Cal Furnace Quick QC Check Pro Cal Cool Muffle Pro Cal Soaked Muffle Pro Cal Furnace Connect Instrument Box Pro Cal Furnace Ini...

Страница 10: ...ce temperature to go very high Remove the Pro Cal Kit and verify the furnace Actual Temperature is tracking the furnace Target Temperature If the furnace itself is not the issue contact Whip Mix Customer Service 102 The difference between the furnace temperature reading and the Pro Cal temperature reading is too great Make sure the Pro Cal tray is inside the furnace and check the cable connections...

Страница 11: ...30 minutes to an hour you can use the Quick Check Selection on the Pro Cal Menu and enter 850 1000 C as a target temperature then re start the Pro Cal cycle afterwords Third use the Quick Check mode and set an 850 C Target Temperature and wait for the furnace Actual Temperature to rise to 850 C Observe the Actual Temperature for a couple of minutes and if the Actual Temperature on the furnace disp...

Страница 12: ...under test via Data Cable Fuse None draws power from furnace under test Data Cable 9 pin DSub Male 9 Pin DSub Female shielded Standard RS232 Computer Cable Maximum Temperature 125 Degrees F System Accuracy 0 125 of measured temperature Calibration Results Accuracy 3 degrees Celsius from 700 C to 1200 C Pollution Category 1 For Indoor Use Only Environmental Ratings Altitude Up to 2 000 meters Ambie...

Отзывы: